X-SERIES CHAIR WILL NOT TILT OR ONLY TILT IN ONE DIRECTION 

1.

 

The tilt feature on X-series chairs is limited to a certain amount of tilt forward and backward 
depending on other included features.  

2.

 

The amount of tilt is controlled by limiting switches behind a small cover on the left side (patient’s 
left) of the chair under the seat section. See Picture below 

3.

 

One of these limit switches may not be working properly. If one of the switches is fixed in the actuated 
position, the tilt motor will not move in that direction.  

4.

 

Using a Phillips head screwdriver, remove the cover and attempt to manually actuate the switches. 

5.

 

An audible click should be able to be heard when the switch is activated and released.  

6.

 

If the one of the switches is not de-activating (button is not “popping” back up), then the switch is 
faulty and needs to be replaced.  

7.

 

Call technical service for part number and price.  

WARNING!

 – THE DEGREE OF TILT IS SET AT THE FACTORY AND SHOULD NOT MODIFIED. 

MODIFYING THE AMOUNT OF TILT MAY CAUSE INTERFERENCE BETWEEN COMPONENTS ON THE 
CHAIR CAUSING DAMAGE TO THE COMPONENTS OR THE DEVICE.   

You may now continue troubleshooting your device or return to the 

General Issues

 page.

 

 

 

Summary of Contents for TMM Series

Page 1: ...hair s serial number these represent the date the chair was manufactured If you are utilizing an electronic copy of this document words that are bold blue and underlined represent links to other pages...

Page 2: ...g Page 5 a Follow this link if one or more portions of the device is working but some are not moving b If the Stretcher chair is an X Series chair the manufacture date is prior to February of 2012 and...

Page 3: ...esting protocol can be found by following this link B Battery Testing Protocol For A AC Powered chairs ensure the device is plugged into a working 120V outlet AC powered chairs require working batteri...

Page 4: ...eshoot the issue Both devices must have the same or similar control boxes 5 If switching pendants resolved the issue contact our Customer Service Department for a replacement pendant 6 If another pend...

Page 5: ...ely failed and may need to be replaced If this is an X Series chair built prior to February of 2012 additional troubleshooting can be done by following this link X Series Chair built prior to February...

Page 6: ...echnical service for part number and price 7 If the same section of the column is still not working and the chair has a KOM control box then the issue is likely in the column Call or email technical s...

Page 7: ...mit switch appears to be working properly access the control box in the base of the chair and unplug the cord in port 4 WARNING This will disable the limit switch allowing the chair to be lowered whil...

Page 8: ...d in the actuated position the tilt motor will not move in that direction 4 Using a Phillips head screwdriver remove the cover and attempt to manually actuate the switches 5 An audible click should be...

Page 9: ...etermine the faulty part in the chair 3 Note which motor is currently functioning and which is not currently functioning 4 Unplug the motor that is not functioning and plug this motor into the port fo...

Page 10: ...supply B Chairs Only Power supply A Chairs Only Port 10 Pendant pendant extension on X series chair Port 7 Limit Switch TMM3 Only Control Box Port Diagrams KOM control box diagram on next page VCU Co...

Page 11: ...of 2012 Return to one of the following sections Tall T Chair is Not Going to Full Height One or More Functions Have Stopped Functioning or the General Issues page Pendant Port 1 Column Port 2 Back Mot...

Page 12: ...Please refer to section 6 of Winco Mfg LLC s Terms Conditions for additional information concerning our return policy When returning parts please contact our Customer Service Department customerservi...

Reviews: