PARAMETER

DESCRIPTION

CODE/VALUE

F1
Graduations

Specifies number of the Weighingblocks graduations. Value should be consistent
with legal requirements and environmental limits on the useful system resolution.

500 1,000 1,500 2,000 2,500 3,000
4,000, 5,000 6,000 8,000 10,000
12,000 13,000 15,000 20,000 30,000
40,000 50,000

F2
Span Gain

Span Gain is related to A/D integration time. The larger the span gain, the higher the
internal resolution, but the slower the update speed.

25 50 75 100 150 200

F3
Zero Track Band

Selects the range within which the scale will automatically zero. Note that the scale
must be in standstill to automatically zero. Selections are in Display Divisions.

d

= graduation

0d 0.5d 1d 3d 5d

F4
Zero Range

Selects the range within which the scale may be zeroed. Note that the indicator must
be in standstill to zero the scale.

100%

1.9%

2%

F5
Motion Band

Sets the level at which motion is detected by comparing the present display update
with the previous one. If motion is not detected for two seconds or more, scale is in
standstill and can process a Print or Zero command. Maximum value varies
depending on local regulations.

0.25

1d

3d

5d

10d

F6
Digital Filter

Averages weight readings to produce higher accuracy. The higher the filter number,
the greater the accuracy but the slower the response time. Choose 4 or 8 unless a very
fast response is needed.

1

2

4

8

F7
Overload Limit

Selects the desired formula which determines the point at which the indicator shows
overload ("

oooo

"). All selections are based on the primary unit selected in F8.

"FS" = Full scale in primary units.

FS

FS + 2%

FS + 1d

FS + 9d

F8
Calib. Unit

Selects the primary base unit to be used in the calibration process. Also the default
unit for normal operation.

"1"

=

primary unit is lb.

"2" = primary unit is in kg.

1
2

F9
Display Div.

Determines the desired weight increments. Value should be consistent with legal
requirements.

1
2
5

F10
Decimal Pt.

Determines location of the decimal point.

0
0.0

0.00

0.000

0.0000

F11
Batching function.

Activating Batching function (not be used with

EC

-Scales

0 = OFF
1 = ON

F16
Zero Calibration

Places indicator into the zero calibration routine. Scrolling down with the ZERO key
one level begins the procedure.

Press ZERO key to begin sequence

F17
Span Calibration

Places indicator into the span calibration routine. Scrolling down with the ZERO key
one level begins the procedure.

Press ZERO key to begin sequence

F18

F19

F20

F21
Factory Reset

This sub-menu will reset all parameters in the

F

and

A

menu to the default

settings. USE WITH CAUTION!

Press the ZERO key twice to execute.

A1

A2

A3

A4

A5
Disable the lb/kg
Key

Allows the lb/kg key to be disabled and allows use of extended graduation.
"0" = Disable the lb/kg key

"1" = Enable the lb/kg key

2

Activate the extended graduation for VB2-100-EC

0
1
2

Summary of Contents for WEIGHBLOK

Page 1: ...Revision 0502 4 2 03 2001 2002 Transcell Technology Inc Contents subject to change without notice Transcell Technology Inc 975 Deerfield Parkway Buffalo Grove IL 60089 Tel 847 419 9180 Fax 847 419 15...

Page 2: ...Manual Normal Calibration Calibration switch ON PRINT NET GROSS OFF ZERO TARE Communica tions port External power 9 VDC Alkaline batteries The Weighingblock is a robust Scale It is designed for portab...

Page 3: ...enu while a few other parameters are altered through the User menu The configuration section of the manual explains how to use the five front panel keys to maneuver and save settings in both menus The...

Page 4: ...ve to the selection value press ZERO once 7 Increase and decrease the value with PRINT and TARE 8 To save the value press NET 9 Press DOT key to go back to parameter number 10 To go back to normal men...

Page 5: ...accuracy The higher the filter number the greater the accuracy but the slower the response time Choose 4 or 8 unless a very fast response is needed 1 2 4 8 F7 Overload Limit Selects the desired formul...

Page 6: ...field Parkway 800 503 9180 www transcell net Buffalo Grove IL 60089 847 419 1515 transcell transcell net F6 F7 F8 F9 A10 Auto Power Off Period Selects the auto power off time period in minutes Off Dis...

Page 7: ...omplete indicator calibration To adjust the max point using the F17 span adjusts procedure 1 Scroll to F 17 then press ZERO to enter span calibration menu 2 The display will momentarily show C 1 for t...

Page 8: ...erating Mode Indicates a low battery condition Re charge the battery or replace if alkaline batteries Err 0 Span Calibration Mode F17 Keyed in weight value is larger than full scale capacity Use a sma...

Page 9: ...normal mode Loadcell Connections Check the Loadcell with a universal voltmeter see table The left pin is number 1 Terminal J1 Description Colours Loadcell AG Colours Loadcell 1042 E 1 Out 5 VDC Brown...

Page 10: ...appears Basic scale use 1 Press ON to start the scale 2 If necessary press ZERO to obtain a weight reading of zero 3 Press OFF to switch off the scale Hold 3 seconds 4 The scale switches off automatic...

Page 11: ...when filling or discharging One time set up F11 must be 1 and A5 must be 2 normally done when delivered RS 232 PIN NUMBER PIN NAME 2 RXD 3 TXD 5 GND 7 TARGET TTL 5V OV 8 TTL 5V Set the target weight...

Reviews: