25
Restore Factory Calibration (B6) – wireless systems only
Caution
: This action cannot be undone.
1. Switch off the RF digital indicator by pressing and holding the ZERO/OFF key for about 5
seconds.
2. Enter the Setup mode by pressing and holding the ON/PRINT key for about 20 seconds or
until the screen shows “F”.
3. Use the ON/PRINT key to change the display to “B”.
4. Scroll down once using the ZERO/OFF key to enter the “Bluetooth” menu. Scale shows
“B 1”.
5. While in the Bluetooth mode, scroll to "
B 6”.
6. Press the ZERO/OFF key once; the display says “F-CAL”.
7. To view all of the factory corner calibration values, press the ZERO/OFF key once; the indi-
cator automatically displays the calibration values for each corner (1-4) in sequence and
then returns to the “F-CAL” level.
8. To view all of the factory span calibration values, press the UNITS key once; the indicator
automatically displays the following in sequence and then returns to the “F-CAL” level.
Span Point #1 Weight Value
Span Point #1 A/D Count
Span Point #2 Weight Value
Span Point #2 A/D Count
Span Point #3 Weight Value
Span Point #3 A/D Count
9. To exit without restoring the factory calibration values, press the PRINT key; the indicator
revert backs up to B6.
10. To restore the factory calibration values, press and hold the NET/GROSS key for 3 sec-
onds; the indicator briefly displays “dOnE” and then reverts back up to B6.
11. In order to save all parameters, it is imperative to exit all menus in the following manner: At
the parameter level, press the UNITS (up) key twice (two times) to exit the Bluetooth Menu.
The indicator displays ‘SAvE’ and then automatically powers off.