24
Key-In Zero Calibration Value (F19)
Note:
This procedure is intended for emergency use only in the case of non-volatile memory
loss. A valid zero calibration value, obtained from a successful F16 calibration procedure,
must be used.
1.
While in the Setup mode, scroll to "
F 19
", and then scroll down once using the ZERO key.
The display will momentarily show "
CAL 0
", then “
E CAL0
” followed by a value of zero
2.
Use the four directional keys to enter in the actual zero calibration value.
3.
After entering the exact value, press the NET/GROSS key to save the value. The display
will revert back up to F19.
Key-In Span Calibration Value (F20)
Note:
This procedure is intended for emergency use only in the case of non-volatile memory
loss. Valid span calibration values, obtained from a successful F17 calibration procedure,
must be used.
1.
While in the Setup mode, scroll to "
F 20
", and then scroll down once using the ZERO
key. The indicator will prompt you to enter the information in the Table below.
2.
If the value shown is correct, press the NET/GROSS key to move to the next parameter.
Otherwise, use the four directional keys to enter in the actual calibration value
3.
After setting the exact value, press the NET/GROSS key to save the value.
4.
At the completion of the sequence, the indicator will then revert back up to F20.
CODE
NAME
ET T 1
First Test Weight Value
ET C 1
First Span Calibration Value
ET T 2
Second Test Weight Value
ET C 2
Second Span Calibration Value
ET T 3
Third Test Weight Value
ET C 3
Third Span Calibration Value
Calibration Value Entry Table
In order to save all parameter settings, it is imperative to exit all menus in the following manner:
At the parameter level, press the UNITS (up) key twice (two times) to exit the Setup Menu. The
indicator displays ‘SAvE’ and then automatically powers off.