Operation Manual for Hand-held Refractometer for 
Specific Gravity & Salinity measurement

 

 
 

 
 
Product Layout

 

 
 
 
 
 
 

 
 
 

 
 
Specification: 

 

RSA0100 
Unit of measurement: 

Specific Gravity  /  Parts-per-thousand 

Measuring range: 

1.000 to 1.070 sg  /  0 to 100ppt 

Display Resolution: 

0.001 sg                / 1ppt 

 

RSA0028 

Unit of measurement: 

Percentage of salt

 

Measuring range: 

0 to 28% 

Display Resolution: 

0.2% 

 
Temperature Compensation: 

Automatic from 10 to 30degree Celsius.  

(ATC model only) 
 
 

Calibration:

 

 

1. 

Calibration should be made in an ambient environment of 20degree Celsius. 

2. 

Open the Light Plate & place few drops of distilled water ½ fill the prism. 

3. 

Close the Light Plate.  Ensure the water is spread across the entire surface of 
the Prism.  There must not be any bubbles or dry spots on the Prism, which is 
cause by too little liquid use. 

4. 

Wait  for  about  30  seconds,  this  is  to  allow  the  temperature  compensation  to 
take place. 

5. 

Hold the meter in the direction of light source and look into the Eyepiece. 

6. 

Adjust the Focus Dial until image is clear and sharp.  Each person has his own 
focusing adjustment. 

7. 

Remove the rubber cap on the Calibration knob, with the use of the provided 
trimming tool and adjust it until the bottom line of the blue section is at 1.000sg 
or 0%. 

8. 

Calibration Completed. 

9. 

Wipe  dries  the  Light  Plate  and  Prism  with  soft  lint-free  clothes  before 
proceeding with measurement. 

 

Operation Manual for Hand-held Refractometer for 
Specific Gravity & Salinity measurement

 

 
 
 
 

Product Layout

 

 
 
 
 
 
 

 
 
 

 
 
Specification: 

 

RSA0100 
Unit of measurement: 

Specific Gravity  /  Parts-per-thousand 

Measuring range: 

1.000 to 1.070 sg  /  0 to 100ppt 

Display Resolution: 

0.001 sg                / 1ppt 

 

RSA0028 

Unit of measurement: 

Percentage of salt 

Measuring range: 

0 to 28% 

Display Resolution: 

0.2% 

 
 
Temperature Compensation: 

Automatic from 10 to 30degree Celsius.  

(ATC model only) 
 

 
Calibration:

 

 

1. 

Calibration should be made in an ambient environment of 20degree Celsius. 

2. 

Open the Light Plate & place few drops of distilled water 1.2 fill the prism. 

3. 

Close the Light Plate.  Ensure the water is spread across the entire surface of 
the Prism.  There must not be any bubbles or dry spots on the Prism, which is 
caused by too little liquid use. 

4. 

Wait  for  about  30  seconds,  this  is  to  allow  the  temperature  compensation  to 
take place. 

5. 

Hold the meter in the direction of light source and look into the Eyepiece. 

6. 

Adjust the Focus Dial until image is clear and sharp.  Each person has his own 
focusing adjustment. 

7. 

Remove the rubber cap on the Calibration knob, with the use of the provided 
trimming tool and adjust it until the bottom line of the blue section is at 1.000sg 
or 0%. 

8. 

Calibration Completed 

9. 

Wipe  dries  the  Light  Plate  and  Prism  with  soft  lint-free  clothes  before 
proceeding with measurement. 

 

Eyepiece 

Calibration Knob 

Focus Dial 

Light Plate 

Prism 

Eyepiece 

Calibration Knob 

Focus Dial 

Light Plate 

Prism 