4.11 Configuration and Calibration Command List
Following is a list of commands, which will operate in the
ADJ
and
TCICAL
mode.
Top-Level Menu Commands
Action
ADJ <enter>
Enter
Adjustment
mode.
TCICAL<enter>
Enter
TCICAL
mode.
Q
Quit
and resume normal operation.
Sub-Menu (ADJ) Commands
Action
CLR <enter>
Clear
all EEPROM data.
D <enter>
Display
to list all EEPROM error correction table data.
P <enter>
Displays current serial
Port
settings, see
§4.6
Serial Port Software Configuration.
Pabc<enter>
Port
protocol assign in
ADJ
mode, see
§4.6 Serial
Port Software Configuration.
Pab<enter>
Port
protocol assign in
TCICAL
mode. Factory set to
P12
. Set to
P22
for One-Foot resolution data on TxD4
& TxD5.
See
§4.7.
Q
Quit
and return to top-level commands.
Saa<enter>
Set
digitizer to 1K altitude (
aa
) mark at current input
pressure. See
§4.8 Dynamic Calibration Adjustment
Procedure.
NOTES:
1. Backspace does not function. If a typing error occurs hit
Q
and begin again.
2. Altitude values 90 to 99 in EEPROM contain factory calibration data and are not customer
accessible.
3. A maximum error of ±499 feet may be introduced at any one altitude.
4.
CLR
clears
all
error data in the EEPROM, and returns the digitizer to the original factory calibration.
5.
ERR
indicates a syntax error.
Summary of Contents for SSD120-N-RS5
Page 17: ...Figure 1 Altitude Digitizer Correspondence Page 17 of 57...
Page 19: ...Figure 2 Dynamic Calibration Set Up Using the ECP 100 Page 19 of 57...
Page 50: ...Wiring Harness Diagram Part Number 881404 50...
Page 51: ...Span Adjust Block Diagram 51...
Page 52: ...Dynamic Calibration Block Diagram 52...
Page 53: ...Figure 2 Temperature vs Warm up Time TEMPERATURE Degrees Celsius Seconds 20 TIME 53...
Page 54: ...Outline Drawing 54...