TrackSense
®
LyoPro
Version 3.3
Page 31 of 53
User Manual
User Manual TrackSense LyoPro – MAN 7
8.2.1.
Ellab Temperature Standard (ETS)
The ETS can be operated as an integrated part of the ValSuite software. It is a
temperature standard that can be started, stopped and used in report
analysis and calibrations.
8.2.1.1.
ETS Versions and Measuring Ranges
*Accuracy between -50 and +150 °C
8.2.1.2.
Proper Use
•
Although the ETS is quite robust, it should be handled with
care. Dropping or causing mechanical shock will loosen or
damage the electronic connections. The platinum resistor is
extremely sensitive, drift can therefore occur if not handled
properly
•
Only use cables (USB) provided by Ellab
•
The ETS system must be properly grounded
•
Never use the ETS outside the measuring range
•
Any ETS with recurring problems should be returned to Ellab
for service
•
In order to maintain the accuracy of the ETS, it is
recommended that they are recalibrated by Ellab at least
once every 12 months
Ordering
code
Type
Temp. Range
Accuracy
20000190 ETS10 -80 to +250 °C
±0.025 °C
±0.015 °C*
20000191 ETS20 -50 to +150 °C
±0.020
20000192 ETS25 -196 to +420 °C
±0.025
20000193
ETS-
A182
-45 to + 150 °C
± 0.025 °C
20000194
ETS-
A213
-95 to +150 °C
± 0.025 °C