TrackSense
®
LyoPro
Version 3.3
Page 27 of 53
User Manual
User Manual TrackSense LyoPro – MAN 7
6.1.8.1.
Proper Use
•
Never use the sensor outside the measuring range
•
Only use the special screwdriver provided by Ellab to remove
sensor from the data logger. Never use any other tools as this
can damage the screw terminals
•
Any data logger/sensor with recurring problems should be
returned to Ellab for service
•
To maintain the accuracy of the TrackSense LyoPro data logger,
it is recommended that it is re-calibrated by Ellab at least once
every 12 months
•
Test all new sensors before using them in any application
•
Always hold the data logger/sensor by the data logger and not
the sensor
7.
Battery
7.1.
TrackSense
®
LyoPro Battery
Ellab offers a special battery with a green
label, which should exclusively be
used for lyophilization applications. This battery is by performance identical to
the standard red and blue batteries, but as label is different (shorter) only this
battery will work in the LyoPro logger. At the same time, the different color
helps identify the correct battery.
Users will have to manually keep track of the number of hours each battery has
been used and the date it was inserted into data logger (use Staedtler Pen). In
the battery time manager, 100% of the battery power would be equivalent to
the default values in ValSuite. For example, the expected lifetime of the
LyoPro battery is 27pprox.. 2,400 hours in applications at -60 °C with a 30
second sample rate and 27pprox.. 1,500 hours at -60 °C with a 10 second
sample rate.
Ordering
code
Type
Temp. Range
Expected
lifetime
Battery
type
Shelf
time
31200050
TSL-BAT-150
with Activator
-80 to +150 °C*
2,400 hours*
Lithium
10 years
31200051
TSL-BAT-150
without
activator
-80 to +150 °C*
2,400 hours*
Lithium
10 years