TrackSense
®
LyoPro
Version 3.3
Page 23 of 53
User Manual
User Manual TrackSense LyoPro – MAN 7
Reading TrackSense LyoPro Data Loggers and Creating Reports:
12.
Place the data loggers into the Reader Station. Go to
LyoPro
Read Logger
,
then
Select All
and
Scan
13.
Select
Read
.
The software will read all data loggers and display the data in the
graph. Select
Close
14.
From the main menu, choose
Session Add
to select the appropriate
Reports
and
Data Analyzes
15.
To print a report, click
File
. A Print window will appear with various
options. Select the desired modules and click
OK
Graph Features Review:
Right click the mouse when it is placed on the graph:
A. Select
Data Viewer
– compare data sets
B. Select
Interval Data Viewer
– numerically display data
C. Select
Add Label
– place special information directly in the graph
D. Select
Copy Graph
– copy graph to the clipboard
E. Select
Set Axis Min/Max
– scale a graph either automatically or fixed
F. Select
Properties
– define data series
For more information on the many other ValSuite software features, please consult
the separate manual delivered with the system.
6.
TrackSense
®
LyoPro Loggers
6.1.
TrackSense
®
LyoPro
6.1.1.
Data Logger Specifications
Ordering
codes
Logger type
Temp Range
Operating
Pressure
Data
Points
Dimensions
Min / Max.
Sample rate
Sensor
31201000
Measuring Range:
-62 to +62 °C
Calibrated Range:
-60 to 60 °C
0.001 mBar
to 4 Bar abs.
100,000
2x
Ø18x32mm
Min. 1 sec.
Max. 23:59
(Hours: Min)
Thermocouple
mounted by
screw
terminal