1.
Ensure that the battery has enough capacity to cover the expected session
length. If the battery is new, activate it separately in the delivered activator for
60 minutes before placing it in the LyoPro data logger with the
+
pole facing
downwards
2.
If not already mounted, connect the thermocouple sensor to the data logger
using the supplied screwdriver. It is important that the
RED
wire goes in the
RED
screw terminal. And that the
BLUE
wire goes in the
BLUE
terminal. Do not
tighten too hard in order to avoid possible damage to the thermocouple
3.
Place the data logger(s) in the reader station slot(s)
4.
Select
LyoPro
Hardware Settings
in ValSuite main menu and chose
USB
(default) or
Ethernet connection
and click
OK