Trace2O HydroTest HTCOD-TR Instruction Manual Download Page 9

 

 

5. 

Maintenance

 

The unit is protected by  a 4AT fuse. The position  of the fuse holder is  on the  reverse 
side  of  the  unit  under  the  main  switch. 
Should the fuse need changing, disconnect the unit from the power supply and open 
the cover with a suitable tool to access the fuse. 

 

 

6. 

Cleaning

 

No special maintenance is necessary apart from periodic cleaning of the unit. 
Disconnect the unit from power supply and use a dust-free cloth with a 
non flammable, non aggressive detergent to clean the unit. 
ATTENTION: 
If  the reactor is contaminated by spillage of the tube contents or breakage of the  
test tube, the disposal of waste (both glass and liquid) must be done according to the 
instructions set out in the Material Safety Data Sheet (MSDS) (Chapter 6 and/or 13). 
A contaminated aluminium block must be replaced prior to further use of the reactor. 
The reactor should be sent to the manufacturer or an authorised service centre. 

 

 

7.

 

Technical  data

 

Power supply

 

 
 
 

V/Hz 

 
 
 

230  /  50-60  or 

 

 

 

115  /  50-60  selectable 

 

Power

 

550 

 

Size

 

mm 

248 x 219 x 171 

 

Weight

 

kg 

3.9 

 

Construction  materials

 

 

Housing: 

ABS 

 

 

 

Protection grid:    PPS 

 

 

 

Lid: 

PC 

 

 

 

Block  insert: 

PBT 

 

 

 

Heating block:     Aluminium 

 

Holes in the

 

 

 

 

aluminium block 

 

24 holes, ø 16.2 mm ± 0.2 mm 

 

Selectable   temperatures

 

°C 

100  /  120  /150 

 

Probe type

 

 

Pt100 A class 

 

Temperature stability

 

°C 

± 1 

 

at the Pt100

 

 

 

 

Selected time 
Heating up from

 

min 

30 / 60 / 120 / continuous (

 

(20°C --> 150°C) 

min 

12 

 

Thermoregulation

 

 

Microprocessor 

 

Protection

 

°C 

at the Alublock for 190 

 

against overheating

 

 

 

 

Beeper

 

dB 

max. 88 

Environmental conditions (operation)

 

Temperature

 

°C 

10 – 40 

Humidity

 

max. 85 

Summary of Contents for HydroTest HTCOD-TR

Page 1: ...HTCOD TR HydroTest COD Thermoreactor Instruction Manual...

Page 2: ...symbols 3 1 2 Important information 4 1 3 Unpacking 4 1 4 Connecting 4 1 5 Buttons 6 1 6 Function of buttons 6 2 Work session 19 3 Beeper 19 4 Function schematics 20 5 Maintenance 21 6 Cleaning 21 7...

Page 3: ...and specified in the corresponding method descriptions Don t exceed temperatures or time spans in any case All warning labels must NOT be removed and should be replaced if they become damaged or fade...

Page 4: ...the ventilation slots Air circulation is necessary to prevent the reactor from overheating If the air circulation is restricted it could cause fire or damage the reactor Never open the reactor housin...

Page 5: ...he packing Part list Part Quantity 1 Thermoreactor HTCOD TR 1 2 Power cord European version 1 3 Instruction manual 1 1 4 Connecting On the reverse side of the reactor Selector for voltage 115 V 230 V...

Page 6: ...matically switches off Heat key By pressing this key after switching the instrument on using the main on off switch see page 17 the reactor will heat up to the pre selected temperature Temp key By pre...

Page 7: ...nently By pressing the Start key the timer starts the work cycles corresponding to the cho sen time span indication by Time LED Starting the work cycle the Time LED changes from lighted to flashing Wh...

Page 8: ...aiting for pressing START key On On On Press Status 1 Press for t 30 60 120 Selection possible Temperature controlled for stability of the selected temperature After pressing Start key Status 4 t Igno...

Page 9: ...nstructions set out in the Material Safety Data Sheet MSDS Chapter 6 and or 13 A contaminated aluminium block must be replaced prior to further use of the reactor The reactor should be sent to the man...

Page 10: ...10 8 Wiring diagramm...

Page 11: ...ff off on on yes 4 Wiring problem internal connection incomplete on off on off off yes 5 Heating problem no power reactor power probe problem on off on off on yes 6 T value underrange Mainboardfaulty...

Page 12: ...12 Trace2o Limited The Technology Centre Wagtech Court Station Road Thatcham Berkshire RG19 4HZ Tel 44 0 1635 566772 Fax 44 0 1635 873509 sales Trace2o com www Trace2o com...

Reviews: