2. 

Work session

 

This section describes the use of the reactor for a standard application. 
For further applications please refer to “Function schematics” (page 20). 

 

 

After  the  unit  is  switched  on  (main  switch,  reverse  side,  position  I)  the  keypad 
automatically  shows  the  last  selected  temperature  and  time  span. 
The corresponding LEDs are lighted. 

After switching on the block heating does not commence automatically. 

For heating up the unit press the “Heat” key. 
After pressing this key the Heat LED is lighted. 
Before  and  after  pressing  the  “Heat”  key  temperature  and  time  span  still  can  be 
changed. 

During  heating  up  the  Temperature  LED  will  light  intermediately,  when  the  selected 
temperature  is  reached  the  Temperature  LED  is  lighted  permanently. 

By pressing the “Start” key the timer starts the work cycles corresponding to the cho- 
sen time span (indication by Time LED). Starting the work cycle the Time LED changes 
from lighted to flashing. 

When  the  work  cycle  ends  Time  LED  and  Temperature  LED  are  lighted  while  the  Heat 
LED  is  off. 

At the end of the work cycle the heater is switched off. 

 

 
 

3.

 

Beeper

 

Select temperature or time: 

 
 

 

short double beep 

 

 

(two  frequencies) 

 

Switch heater on: 

long beep (one frequency) 

 

Switch heater off: 

long beep (one frequency) 

 

Temperature  reaches the  selected value: 

8 x short  beep 
(two  frequencies) 

 

Start of countdown: 

long beep (one frequency) 

 

End of countdown: 

16  x  short  beep 
(two  frequencies) 

 

Pressing  a  key  which  is  inactive 
(at the  moment): 

short  beep  
(one  frequency) 

 

Malfunction: 

Continuous  beep  (2  frequencies) 
until the instruments is switched 
off with the main switch (position “O”). 
In this case the LED combination 
according pages 12/13 allows a 
failure defi 

Summary of Contents for HydroTest HTCOD-TR

Page 1: ...HTCOD TR HydroTest COD Thermoreactor Instruction Manual...

Page 2: ...symbols 3 1 2 Important information 4 1 3 Unpacking 4 1 4 Connecting 4 1 5 Buttons 6 1 6 Function of buttons 6 2 Work session 19 3 Beeper 19 4 Function schematics 20 5 Maintenance 21 6 Cleaning 21 7...

Page 3: ...and specified in the corresponding method descriptions Don t exceed temperatures or time spans in any case All warning labels must NOT be removed and should be replaced if they become damaged or fade...

Page 4: ...the ventilation slots Air circulation is necessary to prevent the reactor from overheating If the air circulation is restricted it could cause fire or damage the reactor Never open the reactor housin...

Page 5: ...he packing Part list Part Quantity 1 Thermoreactor HTCOD TR 1 2 Power cord European version 1 3 Instruction manual 1 1 4 Connecting On the reverse side of the reactor Selector for voltage 115 V 230 V...

Page 6: ...matically switches off Heat key By pressing this key after switching the instrument on using the main on off switch see page 17 the reactor will heat up to the pre selected temperature Temp key By pre...

Page 7: ...nently By pressing the Start key the timer starts the work cycles corresponding to the cho sen time span indication by Time LED Starting the work cycle the Time LED changes from lighted to flashing Wh...

Page 8: ...aiting for pressing START key On On On Press Status 1 Press for t 30 60 120 Selection possible Temperature controlled for stability of the selected temperature After pressing Start key Status 4 t Igno...

Page 9: ...nstructions set out in the Material Safety Data Sheet MSDS Chapter 6 and or 13 A contaminated aluminium block must be replaced prior to further use of the reactor The reactor should be sent to the man...

Page 10: ...10 8 Wiring diagramm...

Page 11: ...ff off on on yes 4 Wiring problem internal connection incomplete on off on off off yes 5 Heating problem no power reactor power probe problem on off on off on yes 6 T value underrange Mainboardfaulty...

Page 12: ...12 Trace2o Limited The Technology Centre Wagtech Court Station Road Thatcham Berkshire RG19 4HZ Tel 44 0 1635 566772 Fax 44 0 1635 873509 sales Trace2o com www Trace2o com...

Reviews: