1.

 

HTCOD-TR

 

 

 

 

1.1 

Introduction

 

Please read the manual before using the unit in particular taking note of the warning 
symbols listed below. 

The manufacture does not take responsibility for any issues caused by use of the unit 
not in accordance with the instructions laid out in this manual. 

 

1.1.1 

Preface

 

The reactor is only suitable for 16 mm Ø test tubes, closed with a lid. 
The unit has a transparent cover, which has to be closed during the heating process. 

The  required  temperatures  and  the  corresponding  time  periods  are  specific  for  the 
different  test  tube  types  and  specified  in  the  corresponding  method  descriptions. 
Don’t  exceed  temperatures  or  time  spans  in  any  case. 

All warning labels must NOT be removed and should be replaced if they become 
damaged or faded. 

 

 

1.1.2 

Guide to symbols

 

The symbols below are used in this  manual to indicate where there is risk of injury or 
damaging  devices  or  to  indicate  especially  useful  information: 

 

 
 

DANGER! 
Indicates risk of injury. 
When  not  following  instructions,  severe  injury  or  death  may  result. 

 

 

ATTENTION! 
Indicates  possible  damage  to  devices. 
When not following instructions, devices may be heavily damaged. 

 

 
 

IMPORTANT! 
Indicates hints on operation and other useful information. 

 

 
 

ATTENTION! 
Hot surfaces! Do not touch, risk to be badly burned! 

 

 
 

Read all instructions before using the instrument.

 

Summary of Contents for HydroTest HTCOD-TR

Page 1: ...HTCOD TR HydroTest COD Thermoreactor Instruction Manual...

Page 2: ...symbols 3 1 2 Important information 4 1 3 Unpacking 4 1 4 Connecting 4 1 5 Buttons 6 1 6 Function of buttons 6 2 Work session 19 3 Beeper 19 4 Function schematics 20 5 Maintenance 21 6 Cleaning 21 7...

Page 3: ...and specified in the corresponding method descriptions Don t exceed temperatures or time spans in any case All warning labels must NOT be removed and should be replaced if they become damaged or fade...

Page 4: ...the ventilation slots Air circulation is necessary to prevent the reactor from overheating If the air circulation is restricted it could cause fire or damage the reactor Never open the reactor housin...

Page 5: ...he packing Part list Part Quantity 1 Thermoreactor HTCOD TR 1 2 Power cord European version 1 3 Instruction manual 1 1 4 Connecting On the reverse side of the reactor Selector for voltage 115 V 230 V...

Page 6: ...matically switches off Heat key By pressing this key after switching the instrument on using the main on off switch see page 17 the reactor will heat up to the pre selected temperature Temp key By pre...

Page 7: ...nently By pressing the Start key the timer starts the work cycles corresponding to the cho sen time span indication by Time LED Starting the work cycle the Time LED changes from lighted to flashing Wh...

Page 8: ...aiting for pressing START key On On On Press Status 1 Press for t 30 60 120 Selection possible Temperature controlled for stability of the selected temperature After pressing Start key Status 4 t Igno...

Page 9: ...nstructions set out in the Material Safety Data Sheet MSDS Chapter 6 and or 13 A contaminated aluminium block must be replaced prior to further use of the reactor The reactor should be sent to the man...

Page 10: ...10 8 Wiring diagramm...

Page 11: ...ff off on on yes 4 Wiring problem internal connection incomplete on off on off off yes 5 Heating problem no power reactor power probe problem on off on off on yes 6 T value underrange Mainboardfaulty...

Page 12: ...12 Trace2o Limited The Technology Centre Wagtech Court Station Road Thatcham Berkshire RG19 4HZ Tel 44 0 1635 566772 Fax 44 0 1635 873509 sales Trace2o com www Trace2o com...