25 |
9 SCAN
You can use Scan menu to take measurements without logging the results.
Scan without logging
1.
Select Scan in the Main Menu and press OK.
2.
Press the SCAN button briefly to make a single measurement or hold the SCAN button down to take
multiple measurements. The scan results for the activated scan angles are displayed.
The device stops scanning as soon as the scan button is released.
Figure 5-1 Scanning without logging
Note: Data scanned in this way is not stored in a batch. For more information on storing scanned data in
batches, see chapter 6.
Scan angles
Depending on the model of the TQC Sheen Glossmeter you are using, you can take measurements for one,
two or three different scan angles. Different scan angles are used for different surfaces. The 60º angle is the
reference angle and can be used for every surface, from matte to mirror finish. The 20º scan angle achieves
the best results on high-gloss surfaces, whereas the 85º scan angle works best on matte surfaces.
For information on activating and deactivating scan angles, see section 4.1. This section also provides
information on how to set the scanning ranges.