16. Troubleshooting
16.1 General Errors
Error Message
Possible Causes
Remedy
Factory Cal.
Failed
then…
See
Handbook
The EEPROM chip which
contains the factory
calibration information has
failed.
The unit must be returned to
TPS for service.
•
Conductivity and Salinity
readings will be accurate only
if used in same range in
which it was calibrated.
•
Temperature readings may be
up to 10% incorrect.
Memory Failed
Calibration Lost
then…
Initialised
MUST ReCalibrate
User calibration settings
have been lost or corrupted.
Re-calibrate the instrument. A
full 2-point calibration will be
required for Conductivity and
Salinity (see sections 3.2 &
4.2) and a 1 point calibration
for temperature (see section
5.1).
Flashing
symbol.
Battery is below 5.60 volts.
Recharge the battery. Note that
the unit will switch itself off
when the battery falls below
5.00 volts.
Meter displays the
word
OFF
, and
switches off.
Battery is below 5.00 volts.
Recharge the battery.
If this fails, check the charger.
If charger OK, replace the
battery.
Meter will not turn
on.
1.
Battery is exhausted.
2.
Faulty Instrument
Recharge the battery.
If this fails, check the charger.
If charger OK, replace the
battery.
Return to factory for repair.
Battery does not
charge up when
charger is connected.
1.
Faulty battery charger or
faulty battery.
2.
Faulty instrument.
1.
Connect the charger and
switch the power on.
Display the battery volts in
the battery saver menu (see
section 11).
If the battery volts are
increasing then the charger
is OK. If the battery volts
do not increase, then the
charger is faulty.
Replace the charger or the
battery, as required.
2.
Return to factory for repair.