Page 28
10.5 Commands
The following commands can be sent from a PC to the
WP-84
. Note that <cr>
denotes carriage return and <lf> denotes a line feed.
Action
Command
Notes
Request current data
?D<cr>
Returns the current Conductivity/
Salinity, Temperature, date and time
from the
WP-84
. The log number
returned is set to Zero.
Request logged data
?R<cr>
Returns all logged records from the
WP-84
memory. The data ends with the
message
ENDS
<cr>
Erase logged data
?E<cr>
Erases all logged records from the
WP-84
memory. Returns the message
ERASED
<cr> to confirm that the
records have been erased.
Request status
information
?S<cr>
Returns the model name, firmware
version number, instrument serial
number and number of logged readings
in memory,
eg:
WP84
ss
ss
V2.0
ss
S1234
ss
9999
<cr>,
where
ss
are spaces. Note that the
number of logged readings is right-
justified.
Request GLP
information
?G<cr>
Returns all calibration GLP information,
plus the instrument model and current
date (see section 10.6 for data format
and hand-shaking).