pH and mV/ORP Troubleshooting, continued...
Unstable
readings.
1. Electrolyte chamber
needs to be refilled.
2. Reference junction
blocked.
3. Glass bulb not clean.
4. Bubble in glass bulb.
5. Faulty connection to
meter.
6. Reference junction not
immersed.
7. KCl crystals around
reference junction, inside
the electrolyte chamber.
Refill with saturated KCl
filling solution.
Clean reference junction, as
per instructions supplied with
the sensor.
Clean glass bulb as per
instructions supplied with the
sensor.
Flick the sensor to remove
bubble.
Check connectors. Replace
if necessary.
Ensure that the bulb AND
the reference junction are
fully immersed.
Rinse electrolyte chamber
with warm distilled water until
dissolved. Replace
electrolyte.
Inaccurate
readings, even
when calibration is
successful.
Reference junction blocked.
Clean reference junction, as
per instructions supplied with
the sensor.
Displays 7.00 for
all solutions.
Electrical short in connector.
1. Check connector.
Replace if necessary.
2. Replace sensor.
Displays 4-5 pH
for all solutions.
Glass bulb or internal stem
cracked.
Replace sensor.
Display flashes
“ATC” and “LIMIT”
The Temperature is not
within the ATC limits.
Cool/Heat solution before
taking measurements.
19.3 Temperature Troubleshooting
Symptom
Possible Causes
Remedy
Displays “
Man
”
when temperature
sensor is plugged
in.
1. Faulty temperature
sensor.
Fit new temperature sensor,
part number 121249.
Temperature
inaccurate and
cannot be
calibrated.
1. Faulty connector.
2. Faulty temperature
sensor.
Check the connector and
replace if necessary.
Fit new temperature sensor,
part number 121249.