Page 38
17. Initialising the WP-80M
If the calibration settings of the
WP-80M
exceed the allowable limits, the
unit may need to be initialised to factory default values. This action may
be required if the sensor is replaced.
To initialise the
WP-80M
:
1. Switch the
WP-80M
off.
2. Press and hold the
key while switching the
WP-80M
back on.
3. The following messages should be displayed:
Initialized
MUST ReCalibrate
then :
WP80MsV5.2 S1234
Meat pH ORP
o
C
4. The meter then displays pH and Temperature. Note that the decimal
points have been replaced with a “
∗
”, to indicate that the unit requires
recalibration.
5. The following items are reset when the
WP-80M
is initialised :
•
Any logged data is erased.
•
Calibration settings for pH and temperature are re-set to factory
defaults.
•
Extra Data Entry is re-set to ON.
18. Instrument firmware version number.
If you need to phone or fax TPS for any further technical assistance, the
version number of your
WP-80M
firmware may of benefit to us. The
version number is displayed by the
WP-80M
at turn-on.