pH and mV/ORP Troubleshooting, continued...
Inaccurate
readings, even
when calibration is
successful.
Reference junction blocked.
Clean reference junction, as
per instructions supplied with
the electrode.
Displays 7.00 for
all solutions.
Electrical short in connector.
1. Check connector.
Replace if necessary.
2. Replace electrode.
Displays 4-5 pH
for all solutions.
Glass bulb or internal stem
cracked.
Replace electrode.
Display flashes
“ATC” and “LIMIT”
The Temperature is not
within the ATC limits.
Cool/Heat solution before
taking measurements.
18.3 Temperature Troubleshooting
Symptom
Possible Causes
Remedy
Displays “
Man
”
when temperature
probe is plugged
in.
1. Faulty temperature
probe.
Fit new temperature probe,
part number 121247.
Temperature
inaccurate and
cannot be
calibrated.
1. Faulty connector.
2. Faulty temperature
probe.
Check the connector and
replace if necessary.
Fit new temperature probe,
part number 121247.