System Data Display sequence. continued…
Cond. Alarm Margin=0.30mS/cm
↑↑
:Next
Cond. Alarm Delay Time=10 minutes
↓↓
:Back
Cond. High Alarm=2.30mS/cm
↑↑
:Next
Cond. Low Alarm=1.50mS/cm
↓↓
:Back
pH Limit=6.5pH Add if High
↑↑
:Next
pH Sensitivity=Medium
↓↓
:Back
pH ON Time=10 mins
↑↑
:Next
pH OFF Time=5 mins
↓↓
:Back
pH ShutOFF Timer=15 mins
↑↑
:Next
pH Quick Dose Time=5 mins
↓↓
:Back
pH Alarm Margin=0.3pH
↑↑
:Next
pH Alarm Delay Time=10 mins
↓↓
:Back
pH High Alarm=7.0pH
↑↑
:Ends
pH Low Alarm=6.2pH
↓↓
:Back
When the Plus-Pack is fitted, an additional screen showing the currently selected RS232 baud rate
is provided.