26
9.3 pH Calibration Messages
1. If a 1-point calibration has been successfully performed, the
hydro
CHEM
will display the
following message and the asymmetry of the electrode. Note that the slope value from the last
calibration is also shown. For example…
Asymmetry Calibration Successful
+0.10pH Asym 100.0% Slope
2. If a 1-point calibration has failed, the
hydro
CHEM
will display the following message, and the
failed asymmetry value of the electrode. For example…
Calibrate Failed, +1.1 pH Asymmetry
Repeat Cal. or Initialise Calibration
3. If a 2-point calibration has been successfully performed, the
hydro
CHEM
will display the
following message, and the asymmetry and slope of the electrode. For example…
Slope & Asymmetry Calibration Successful
+0.10pH Asym 99.0% Slope
4. If a 2-point calibration has failed, the
hydro
CHEM
will display the following message, and the
failed slope value of the electrode. For example…
Calibrate Failed, 80% Slope
Repeat Cal. or Initialise Calibration