Page 7
1.3 Specifications
Probe Type ...........................High sensitivity 90
O
probe is standard.
Wide range 180
O
retro-scatter probe is optionally
available.
With High Sensitivity 90
O
Probe
Ranges*................................0 to 30.00 NTU
0 to 300.0 NTU
0 to 3,000 NTU
Resolution ........................................0.02 NTU
0.1 NTU
1 NTU
With Wide Range 180
O
Probe
Ranges*................................0 to 30.00 NTU
0 to 300.0 NTU
0 to 3,000 NTU
0 to 30,000 NTU
Resolution ........................................0.02 NTU
0.1 NTU
1 NTU
10 NTU
* Exact ranging subject to probe performance. The
WP-89
will always select
the range that provides the best resolution for the reading.
Repeatability ........................2% ±1 digit on all ranges.
Readout Modes.....................Turbidity (NTU)
Relative Turbidity Reference (NTU)
Relative Turbidity REL (Turbidity – Relative
Turbidity Reference)
Good Laboratory Practices....All readings and calibration constants are stored
together with date and time and can be recalled at
any time.
Memory................................100 readings including date and time.
Automatic Logging...............User-set for one reading every 1 to 90 seconds or
minutes.
RS232 Port (optional) ...........4800 baud, 8 bits, no parity, 1 stop bit, XON/XOFF
Protocol.