Page 15
5. Place the probe into a the second turbidity standard solution (eg 360 NTU).
IMPORTANT :
Leave at least 25mm clearance from the bottom of the
probe to the bottom of the beaker to avoid interference with the reading.
The reading from the probe is displayed. For example…
360.3NTU pnt2/2
F2:grab
When the reading has stabilised, press
to GRAB the turbidity standard
solution reading.
6. The
WP-89
will now allow you to enter the value of the second turbidity
standard solution. For example…
360.3NTU pnt2/2
F2:OK
↑↑↓↓
:set
Press
and
to set the value of the turbidity standard solution.
Press
to calibrate the second calibration point.
Alternatively, press
to quit and retain the current calibration.