PERFORMING TIGHTNESS TEST (cont)
5. Pressure P2 will be
displayed.
Tap the list icon to
access the optional
menu.
Tap on Tightness
Test Setting.
6. Tap the list icon to
access the option-
al menu.
Tap on Zeroing
Sensors
7. The tightness Test
settings menu will
display.
Set the alarm point
for the maximum
allowable drop from
the starting pres-
sure.
Set the stabilization
time. This is the
time prior to the
clock starting to
give the pressure
reading time to
become steady.
Set the test time.
Select whether the
alarm should
vibrate.
Tap Apply and
Close.
8. Tap the + icon to
select the SP620
to be zeroed.
Tap the SP620 to
be zeroed.
Tap the Reset icon
to set the pres-
sure reading to 0
(+/- 0.05 counts)
Tap Close to
return to normal
operation.