19
17
c. Measuring DC Amps
CAUTION!
Do not attempt to make a current measurement with the
test leads connected in parallel with circuit to be tested.
Test leads must be connected in series with the circuit.
WARNING!
Do not attempt to make a current measurement of
circuits with more than 600V present. Instrument
damage and /or personal injury may result.
Instrument set-up:
FUNCTION
BLACK RED
MINIMUM
MAXIMUM
TEST LEAD TEST LEAD
READING
READING
µA
COM
mA
0.1µA
4000µA
mA
COM
mA
10µA
400mA
A
COM
A
1mA
10.00A
Measurement Procedure:
1.
Disconnect power to circuit to be measured.
2.
Plug black test lead into the
COM
input jack.
3.
Plug the red test lead into the
mA
or
A
input jack
depending on the value of current to be measured.
4.
Set the rotary switch to the
µA, mA
or
A
function.
5.
Connect test leads in series to circuit to be measured.
6.
Reconnect power to circuit to be measured.
7.
Read the current on the 440.
!
Optional DMM Functions (refer to page 27)
REL
PEAK
DC
AC
FULL
HOLD
FREQ
FREQ
AUTO
F1
F2
F3
F4
F5
PROG (refer to page 27)
REC
REL%
COMP
EDIT
EXIT
F1
F2
F3
F4
F5
Optional D Mode Functions (refer to page 29)
COMBO METER
SCOPE
TREND
EXIT
F1
F2
F3
F4
F5