Instrument Specifications
Operating Temperature Range
-10° to 50°C (14° to 122°F)
Battery
Rechargeable Ni-MH, >6 Hour Life
Charger Input Voltage
115V or 230V : 50/60 Hz AC
Fuels
Natural Gas, LPG, Light Oil, Heavy Oil
OPT 1 = Bituminous Coal
OPT 2 = Anthrachite Coal
OPT 3 = Coke
OPT 4 = Butane
OPT 5 = Wood (Dry)
OPT 6 = Bagasse
Display
Backlit LCD
Data Storage
50 sets of readings
Data Logging
400 sets of readings on
‘Timed Interval Logging’
Time & Date
24 Hour Real Time Clock
Dimensions
200mm x 90mm x 60mm
Weight
500g
Exhaust
Safety Spigot
Conforms To:
BS7927 (and the draft BS7967)
Standard Accessories
Flue Temperature Probe
A770
‘K’-Type Temperature Air Probe
GK11M
In-Line Water Trap Bowl Filter Complete
A794
Mini Pump Protection Filter Assembly
A763
Battery Charger
A766
Protective Boot
A765
Soft Carrying Case
A768
Exhaust Spigot
A764
Temperature Measurement
Range
-40° to 800° C (-40° to 1472° F)
Resolution
1°C or F
Accuracy (K-type)
±0.3%, ±1°C or ±2°F
D r a f t / P r e s s u r e M e a s u r e m e n t
Selectable Ranges
mBar,kPa, and inH2O
Ranges
-150 mBar to +150 mBar
-15 kPa to +15 kPa
-60 inH2O to +60 in H20
Resolution
0.01mBar
Accuracy
±0/5% fsd
Gross/Net Efficiency
Range
0 to 100%
Resolution
0.1%
Oxygen (02) Measurement
Range
0 to 25%
Resolution
0.1%
Accuracy
±0.3%
Carbon Monoxide (CO) Measurement
Range
0 to 10,000 ppm
Resolution
1 ppm
Accuracy
±5 ppm or 5%, whichever is greater
Carbon Dioxide (CO2)
Range
0 to 25%
Resolution
0.1%
Accuracy
Calculated
Ratio (CO/CO2)
Range
0 to 0.999
Resolution
0.001
Accuracy
Calculated
T P I 7 1 2
“ 5 - i n - 1 ” C o m b u s t i o n A n a l y s e r
Optional Accessories:-
A773 - Oil Filter
A788 - Smoke Test Pump
CK21M - Temperature Pipe Clamps
A767 - IRDA to RS232 Adaptor for PC
A772 - PC Software
A746 - 10 Spare Paper Rolls for IR Printer
Plus Various ‘K’ Type Thermocouple Probe
A740
Infra Red Printer
N O T E :
When selecting
OIL as a fuel be
sure to use the
optional (A773)
oil filter or
readings could
become erratic