1. Introduction

Thank   you   for   purchasing   the   D-­‐650   amalgam   mixer.      This   unit   uses
a   microprocessor   to   control   its   functions.      Please   read   this   manual
carefully   before   you   begin   to   use   the   mixer.

2. Usage

This   D-­‐650   is   used   to   mix   dental   amalgam,   glass   ionomer   and
cement;   dental   capsules.      Please   see   the   manufacturer   instructions
for   trituration   and   time   settings   on   each   dental   material.

3. Technical   data

3.1

 

Electric   safety:   Grade   1,   Type   B   

3.2

 

Power   supply:   AC   220/110   ±22v,   50Hz/60Hz   

3.3

 

Power   consumption:   200VA   

3.4

 

Oscillation   freq.:   4250/min   

3.5

 

Fuse:   2A,   Ø5   x20mm   

3.6

 

Operating   circumstance   

3.6.1

 

Working   temperature:   5’C   ~   40’C   

3.6.2

 

RH:   ≤   80%   

3.7

 

Storage   circumstance   

3.7.1

 

Temperature:   -­‐40’C   ~   +55’C   

3.7.2

 

RH:   ≤   80%   

4. Production   indication

1. Cap

5.

Time   (+)

2. Start   switch

6.

Time   (-­‐)

3. Vibration   Arm

7.

Power   switch

4. LCD   Display

8.

Fuse

5. Operating   method

5.1

 

Choose   a   amalgam   capsule   

5.2

 

Activate   the   capsule,   according   to   manufacturer   
instructions.   

5.3

 

Open   the   transparent   protective   cover   of   the   unit,   and   fix   
the   capsule   on   the   holder   of   the   mixer.      Then   close   the   
cover.      

Do   not   triturate   with   the   cover   open

   

5.4

 

Time   preset   

Press   the   “

Δ”   key   to   set   time   longer,   press   the   “

”   key   to

set   timer   shorter.   

5.5

 

Press   the   “Start”   button,   the   mixer   will   begin   to   triturate,   
the   cycle   will   end   at   the   time   specified.      If   you   wish   to   stop   
the   cycle   before   it   is   completed,   press   the   “Start”   button   
again.   

5.6

 

Once   the   trituration   cycle   is   complete,   open   the   cover,   
remove   the   capsule   and   close   the   cover.   

6. Notice

6.1

Use   only   Earth   grounded   receptacle   (wall   outlet).   

6.2

Before   pressing   the   “Start”   button;   make   sure   the   cover   is   
closed.   

7. Maintenance

Lubricate   the   rotating   parts   periodically.

WARNING

:

   At   the   time   of   un-­‐boxing;   make   sure   to   take   off   the   two

side   screws   used   to   secure   the   triturating   arm.

8. Troubleshooting

Symptom   

Solve   method   

1    The   LCD   has   no   numeric   display   

Check   2A   fuse   

2    The   capsule   holder   swings   loosely   

Check   the   driving   rubber   band   

9. Guarantee

If   the   unit   is   not   working   properly   due   to   product   quality,   it   will   be
repaired   or   replaced   free   of   charge   by   TPC.      This   unit   comes   with   a
1   year   warranty,   which   starts   from   the   date   when   the   unit   was
purchased   from   TPC   or   a   dealer,   Warranty   does   not   cover   fuses   or
negligence   by   the   end   user.      TPC   DOES   NOT   COVER   FREIGHT   FOR
ANY   WARRANTY   ISSUES.