Symbol
Symbol
Symbol
Symbol Definitions
Definitions
Definitions
Definitions
CAP.
CAP.
CAP.
CAP.---- ---- ----
- Next content displayed is Capacity
SPL
SPL
SPL
SPL.Lo
.Lo
.Lo
.Lo
- Sample load weight of low point.
CAL.ON
CAL.ON
CAL.ON
CAL.ON
- Calibration seal switch is in the ON position
SPL
SPL
SPL
SPL.HI
.HI
.HI
.HI
- Sample load weight of high point.
CAL.Px
CAL.Px
CAL.Px
CAL.Px
- Calibration on point(x)
SPL
SPL
SPL
SPL.PWT
.PWT
.PWT
.PWT - Sample goods weight to calculate piece weight
CAL.End
CAL.End
CAL.End
CAL.End - Calibration is finished
INP
INP
INP
INP.PCS
.PCS
.PCS
.PCS
- Input pieces number of weighted goods
COMP
COMP
COMP
COMP
- To go to input COMPARE data mode
SPL
SPL
SPL
SPL.PCT
.PCT
.PCT
.PCT
- Sample goods weight to calculate
HIGH
HIGH
HIGH
HIGH
- To input HIGH limitation data of Comparison
INP
INP
INP
INP.PCT
.PCT
.PCT
.PCT
- Input percentage of weighted goods
LOW
LOW
LOW
LOW
- To input LOW limitation data of Comparison
AC.xxxx
AC.xxxx
AC.xxxx
AC.xxxx
- Accumulation instance is xxxx
PR
PR
PR
PR.TARE
.TARE
.TARE
.TARE - To Preset TARE weight
Troubleshooting
Troubleshooting
Troubleshooting
Troubleshooting
SYMPTOM
SYMPTOM
SYMPTOM
SYMPTOM
PROBABLE
PROBABLE
PROBABLE
PROBABLE CAUSE
CAUSE
CAUSE
CAUSE
REMEDY
REMEDY
REMEDY
REMEDY
0
00
0
﹉﹉﹉﹉
﹉﹉﹉﹉
﹉﹉﹉﹉
﹉﹉﹉﹉
Weight reading exceeds the Power On Zero limit
Make sure scale platform is empty. Perform zero
calibration.
0
00
0
﹍﹍﹍﹍
﹍﹍﹍﹍
﹍﹍﹍﹍
﹍﹍﹍﹍
Weight reading is below the Power On Zero limit
Install platform on scale. Perform zero calibration.
﹉﹉﹉﹉
﹉﹉﹉﹉
﹉﹉﹉﹉
﹉﹉﹉﹉
Weight reading exceeds the overload limit or the weight
value can’t be displayed in the current unit of measure
because it exceeds 6 digits
Reduce load on scale until weight value can be
displayed or use an alternate unit of measure.
﹍﹍﹍﹍
﹍﹍﹍﹍
﹍﹍﹍﹍
﹍﹍﹍﹍
Weight reading below minimum load limit
Install platform on scale. Perform zero calibration.
EEP.E2
EEP.E2
EEP.E2
EEP.E2
USER parameters are not correctly set
Re-set settings in USER mode
CAL.Er
CAL.Er
CAL.Er
CAL.Er
Calibration error. Input data or loaded weight is too
small, too large, unstable, or not linear.
Input correct data, load correct weight onto
platform, or service is required
PWT
PWT
PWT
PWT.ER
.ER
.ER
.ER
Piece weight error. The weight on the platform is too
small (<0.5d) to define a valid reference weight.
Use a greater weight for the sample
PCT
PCT
PCT
PCT.ER
.ER
.ER
.ER
Unit percentage weight error. The percentage weight
(1%, 0.1%, or 0.01%) is too small (<0.5d).
Use a greater weight for the sample
STB.ER
STB.ER
STB.ER
STB.ER
Stabilization time USER-HOLD-STB.TIM is too short, or
Hold range USER-HOLD-HLD.RNG is too small, or other
failure
Set USER-HOLD-STB.TIM longer, or set
USER-HOLD-HLD.RNG larger, or service required
Won’t
Won’t
Won’t
Won’t turn on
turn on
turn on
turn on
Power cord not plugged in or properly connected.
Power outlet not supplying electricity.
Battery discharged. Other failure.
Check power cord connections. Make sure power
cord is plugged into the power outlet. Check power
source. Replace batteries. Or service required.
Unable to
Unable to
Unable to
Unable to zero the
zero the
zero the
zero the
display
display
display
display or will not zero
or will not zero
or will not zero
or will not zero
when turned on
when turned on
when turned on
when turned on
Load on scale exceeds allowable limits.
Load on scale is not stable.
Load cell damage.
Remove load on scale.
Wait for load to become stable.
Service required.
Battery symbol is empty
Battery symbol is empty
Battery symbol is empty
Battery symbol is empty
or Lo.bAt is shown
or Lo.bAt is shown
or Lo.bAt is shown
or Lo.bAt is shown
Batteries are discharged
Replace batteries