4 - 1
Oct. 1995 © TOSHIBA CORP.
1550 TROUBLESHOOTING
4. TROUBLESHOOTING
(1)
C21
Optical system initialization errors
1
Turn the power supply ON/OFF, then visually check that the following units are initialized.
Q1: Does the carriage function? (Follow “YES” if the carriage functions even slightly, or sounds as
if it is trying to function.)
C21
Check if J11, J82 are correctly connected.
Check if fuse on PS-ACC is connected.
Q2: Is carriage initialization normal?
NO
NO
Check item
2
“Carriage Errors”.
YES
Q3: Is the Service Call message displayed after the lens and mirror move?
NO
Check item
3
“Lens and Mirror Errors”.
Q4: Use the
TEST MODE CH
(
0
4
) command code 21 or 22 to move the lens and mirror to the
paper-exit position (enlargement position), then turn the power supply ON and OFF again.
When the power is turned ON again, do the lens and/or mirror make an out-of-step noise at
the home position?
NO
YES
For some reason LNS-SW and MRR-SW are not switched ON/
OFF correctly:
●
Check visually if LNS-SW and MRR-SW can clear the bracket.
●
Check for disconnection or short circuiting around pins J80, J81,
J11, J51, J52, J4.
Issue the
TEST MODE CH
command again and visually check
for faulty conditions.
YES