Topcon DRI OCT Triton Quick Reference Manual Download Page 10

10

Topcon (Great Britain) Medical Limited 

DRI OCT Triton | Quick Reference Guide | 

Anterior Segment Scanning

Select 6 mm Anterior Radial Scan to assess corneal thickness
and instruct the Patient to look straight ahead at the little red
dot in the camera lens.

3

To capture, press the trigger on the joystick.

4

Images will be saved automatically onto Imagenet 6.

5

2. Once the cornea has come 
into view, ensure it is sat in 
the yellow box by pushing 
the machine forwards and 
by twisting the joystick up 
or down to make sure the 
Corneal Reflex line is visible. 

1. Align the radial scan with 
the centre of the pupil and 
drive the machine slowly 
forward until you see the 
cornea in the left hand box

To capture an Anterior Scan, instruct the Patient to look straight ahead and
ensure they are aligned with the callipers on the side of the headrest. Attach the Anterior
Lens and the Anterior Headrest.

1

Select 16 mm or 6 mm Line Anterior to assess the Patient’s angles.
There is no internal fixation, however just instruct Patient to look
straight ahead or turn the External Fixator on.

2

Anterior Segment Scanning

Anterior Headrest

2. Once you see the cornea, 
drive the machine further 
forward until the iris is sat
in between the blue lines 
and press the trigger on 
the joystick to capture.

1. Align the line scan through 
the centre of the pupil and 
drive the machine slowly 
forward until you see the 
cornea in the left hand box.

Anterior Lens

9

Topcon (Great Britain) Medical Limited 

DRI OCT Triton | Quick Reference Guide | 

Anterior Segment Scanning

Anterior Headrest

Anterior Lens

10

61647 Topcon quick start guide Triton.indd   10

27/03/2019   09:56

Summary of Contents for DRI OCT Triton

Page 1: ...Quick Reference Guide DRI OCT Triton Optical Coherence Tomography 61647 Topcon quick start guide Triton indd 1 27 03 2019 09 56...

Page 2: ...to Fluorescence Peripheral Fundus Photography OCT Angiography Fluorescein Angiography Page 3 Page 5 Page 8 Page 9 Page 11 Page 13 Page 15 Page 17 Topcon Great Britain Medical Limited DRI OCT Triton Qu...

Page 3: ...n Medical Limited DRI OCT Triton Quick Reference Guide Select Patient 3 Select Patient In the case of a New Patient Click New Patient and access the register Patient panel Input the necessary items an...

Page 4: ...n Medical Limited DRI OCT Triton Quick Reference Guide Select Patient Select Patient In the case of a New Patient Click New Patient and access the register Patient panel Input the necessary items and...

Page 5: ...al Limited DRI OCT Triton Quick Reference Guide Capturing OCT On Off Position the Patient Choose Scan Type Place the Patient s chin on the chinrest Keep their chin and forehead stable Be sure to adjus...

Page 6: ...are green and overlapping and in the brackets Yellow too close Orange too far away Green line Scan position Blue cross Fixation target position 6 Topcon Great Britain Medical Limited DRI OCT Triton Qu...

Page 7: ...o fixation losses or blinks If poor quality press Delete and repeat capture process 7 Topcon Great Britain Medical Limited DRI OCT Triton Quick Reference Guide Capturing OCT Position the Patient Choos...

Page 8: ...tain Medical Limited DRI OCT Triton Quick Reference Guide Scan Types Scan Types Macula 3D Macula Scans the macula area 7 x 7 mm cube scan 3D Wide Scans macula and disc combined 12 x 9 mm scan FGA Mode...

Page 9: ...k Reference Guide Anterior Segment Scanning Anterior Headrest Anterior Lens To capture an Anterior Scan instruct the Patient to look straight ahead and ensure they are aligned with the callipers on th...

Page 10: ...rior Scan instruct the Patient to look straight ahead and ensure they are aligned with the callipers on the side of the headrest Attach the Anterior Lens and the Anterior Headrest 1 Select 16 mm or 6...

Page 11: ...and Down chinrest buttons found next to the joystick on the machine Topcon Great Britain Medical Limited DRI OCT Triton Quick Reference Guide Fundus Photography Fundus Photography Colour or Auto Fluo...

Page 12: ...please turn up the illumination level Save Topcon Great Britain Medical Limited DRI OCT Triton Quick Reference Guide Fundus Photography Fundus Photography Colour or Auto Fluorescence 1 Position the P...

Page 13: ...p and Down chinrest buttons found next to the joystick on the machine Topcon Great Britain Medical Limited DRI OCT Triton Quick Reference Guide Peripheral Fundus Photography 0 93 M 2 0 538 0 3 9 33 0...

Page 14: ...umination If IR fundus image is noisy or dark please turn up illumination level Topcon Great Britain Medical Limited DRI OCT Triton Quick Reference Guide Peripheral Fundus Photography 0 93 M 2 0 538 0...

Page 15: ...just the Patient s height using the Up and Down chinrest buttons found next to the joystick on the machine en B scan image quality r and in the red in the bove press Optimize to ce signal strength Top...

Page 16: ...strength Topcon Great Britain Medical Limited DRI OCT Triton Quick Reference Guide OCT Angiography ce the Patient is d correctly you can t the OCT scan to e it is centrally in the y moving this Z Tab...

Page 17: ...and Down chinrest buttons found next to the joystick on the machine Fluorescein Angiography 1 Your hospital will have its own protocol on administering fluorescein and timings of the procedure 2 3 Mo...

Page 18: ...sity Increase if the Patient is poorly dilated or has small pupils Fluorescein Angiography 1 Your hospital will have its own protocol on administering fluorescein and timings of the procedure 2 3 Move...

Page 19: ...n co uk w topcon medical co uk Topcon Ireland t 353 1 223 3280 e medical ie topcon com w topcon medical ie Topcon Great Britain Medical Limited Topcon House Kennetside Bone Lane Newbury Berkshire RG14...

Reviews: