58
OBJECTIVE OPERATIONS
9
Using the control lever, move the instrument body and put the alignment bright spot into the ( )
scale.
10
When the positional relation between the instrument body and the patient's eye is proper for pho-
tography after putting the alignment bright spot into the ( ) scale, the fundus photography is auto-
matically done by the auto shoot function. Ask the patient not to blink and not to move his/her
eye during photography.
NOTE
• By this operation, focusing and positioning are performed. However, photo-
graphing is not performed.
• By this operation, the alignment bright sport is moved to the left.
• When the auto focus function is ON, even if you pull the base unit toward the
patient in stereoscopic photography, the focusing knob is not reset to the initial
position.
• When the auto focus function is OFF, turn the focusing knob to change the
split lines into one line.
• When the auto shoot function is OFF or when the diopter compensation lens
is set to others except "0", put the alignment bright spot into the ( ) scale and
then select the
.
NOTE
• If the patient blinks right before the fundus photography, the automatic blink
detection function stops the photography. Start the procedure again from
alignment and take a picture.
• When the auto shoot function is OFF or when the diopter compensation lens
is set to others except "0", the instrument does not take a picture automati-
cally. Press the
to take a color picture of fundus.
STEREOSCOPIC PHOTOGRAPHY SELECTOR BUTTON
Stereoscopic
photography
selector button
PHOTOGRAPHY BUTTON