116
GENERAL INFORMATION ON USAGE AND MAINTENANCE
Guidance and manufacturer's declaration - electromagnetic immunity
The DRI OCT-1 Model Triton is intended for use in the electromagnetic environment specified below. The
customer or the user of the DRI OCT-1 Model Triton should assure that it is used in such an environment.
Immunity test
IEC 60601
test level
Compliance
level
Electromagnetic environment -
guidance
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
3 Vrms
150kHz to 80MHz
3 V/m
80MHz to 2.5GHz
3 V
3 V/m
Portable and mobile RF communications
equipment should be used no closer to any
part of the DRI OCT-1 Model Triton, including
cables, than the recommended separation
distance calculated from the equation appli-
cable to the frequency of the transmitter.
Recommended separation distance
d
= 1.2
d
= 1.2
80MHz to 800MHz
d
= 2.3
800MHz to 2.5GHz
where
P
is the maximum output power rating
of the transmitter in watts (W) according to
the transmitter manufacturer and
d
is the
recommended separation distance in meters
(m).
Field strengths from fixed RF transmitters,
as determined by an electromagnetic site
survey,
a
should be less than the compliance
level in each frequency range.
b
Interference may occur in the vicinity of
equipment marked with the following sym-
bol:
NOTE 1
NOTE 2
At 80 MHz and 800 MHz, the higher frequency range applies.
These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.
a
b
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones
and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be pre-
dicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF trans-
mitters, an electromagnetic site survey should be considered. If the measured field strength in the
location in which the DRI OCT-1 Model Triton is used exceeds the applicable RF compliance level
above, the DRI OCT-1 Model Triton should be observed to verify normal operation. If abnormal per-
formance is observed, additional measures may be necessary, such as reorienting or relocating the
DRI OCT-1 Model Triton.
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
P
P
P