SOL-96.110.081 / 260422
TOPAZ Iron
Troubleshooting
48
Vessel is not emptied and/or the vessel is overflowing
through vent tube
Photometer does not fill
Measured value is not updated
Screen does not work
Possible cause
Corrective action
Tubes are blocked
– Check condition of emptying
tubes (inside and outside).
Emptying tube is pinched
– Make sure that the emptying
tube is not pinched when closing
the door.
Emptying solenoid valve
(SVA) does not switch
– Check that the solenoid valve is
plugged in correctly.
– If necessary, change the
membrane.
Possible cause
Corrective action
Pump does not work
properly
– Check that pump is connected
correctly and that it is not
blocked.
One or more tubes are
blocked or pinched
– Check condition of all sample
tubes.
Possible cause
Corrective action
The same value was
measured twice
– No action necessary.
Measurement was stopped
or paused
– Start measurement again.
Possible cause
Corrective action
The application does not
react.
– Restart the analyzer.
Screen is damaged.
– Contact customer support.