background image

 

 

 

 

 

 

 

 

 

O

O

P

P

E

E

R

R

A

A

T

T

I

I

O

O

N

N

 

 

M

M

A

A

N

N

U

U

A

A

L

L

 

 

 

 

TH2523/A        Battery Tester

 

  

 

  

 

  

 

  

C

C

C

h

h

h

a

aa

n

n

n

g

gg

z

zz

h

h

h

o

oo

u

u

u

 

  

T

T

T

o

oo

n

n

n

g

gg

h

h

h

u

u

u

i

ii

 

  

E

E

E

l

ll

e

ee

c

cc

t

tt

r

rr

o

oo

n

n

n

i

ii

c

cc

 

  

C

C

C

o

oo

.

..

 

  

,

,,

 

  

L

L

L

t

tt

d

dd

.

..

 

  

 

  

 

  

A

A

A

d

dd

d

dd

:

:: No.3, Tianshan Road, New District, Changzhou, Jiangsu

 

T

T

T

e

ee

l

ll

:

::

0

00

5

55

1

11

9

99

8

88

5

55

1

11

3

33

2

22

2

22

2

22

2

22

8

88

5

55

1

11

1

11

3

33

3

33

4

44

2

22

 

  

F

F

F

a

aa

x

xx

:

::

0

00

5

55

1

11

9

99

8

88

5

55

1

11

0

00

9

99

9

99

7

77

2

22

 

  

E

E

E

-

--

m

m

m

a

aa

i

ii

l

ll

:

::

 

  

S

S

S

a

aa

l

ll

e

ee

s

ss

@

@

@

t

tt

o

oo

n

nn

g

gg

h

hh

u

uu

i

ii

.

..

c

cc

o

oo

m

m

m

.

..

c

cc

n

nn

 

  

W

W

W

e

ee

b

bb

s

ss

i

ii

t

tt

e

ee

:

::

 

  

h

hh

t

tt

t

tt

p

pp

:

::

/

//

/

//

 

  

w

w

w

w

w

w

w

w

w

.

..

t

tt

o

oo

n

nn

g

gg

h

hh

u

uu

i

ii

.

..

c

cc

o

oo

m

m

m

.

..

c

cc

n

nn

 

  

 

 

 

Summary of Contents for TH2523/A

Page 1: ...ew District Changzhou Jiangsu T T Te e el ll 0 0 05 5 51 1 19 9 9 8 8 85 5 51 1 13 3 32 2 22 2 22 2 22 2 2 8 8 85 5 51 1 11 1 13 3 33 3 34 4 42 2 2 F F Fa a ax x x 0 0 05 5 51 1 19 9 9 8 8 85 5 51 1 1...

Page 2: ...peration 4 2 1 Introduction to Front Panel 4 2 2 Introduction to Rear Panel 5 2 3 Display Zones 7 2 4 Keys and Their Corresponding Display Pages 8 2 4 1 Main Menu Key DISP 8 2 4 2 System Setup Key SET...

Page 3: ...29 4 1 7 Time 29 4 2 File Manage 29 4 2 1 Introduction to Save Recall 30 4 2 2 Structure of File Folder File in a U Disk 30 Chapter 5 Performance Index 34 5 1 Measurement Function 34 5 1 1 Measurement...

Page 4: ...pter 7 SCPI Command Reference 49 7 1 TH2523 Subsystem Commands 49 7 1 1 DISPlay Subsystem Commands 49 7 1 2 FUNCtion Subsystem Commands 50 7 1 3 APERture Subsystem Commands 57 7 1 4 TRIGger Subsystem...

Page 5: ...is instrument Introductions to the improvements of the instrument in performance function internal structure outer appearance accessories packing material etc are subject to change without notice If y...

Page 6: ...ducts require batteries to work properly quickly and accurately detecting the battery performance will directly affect the product performance With profound electronic technology and extensive market...

Page 7: ...under the temperature ranging from 0 to 40 relative humidity of no greater than 75 For high accuracy use the instrument in the environment above mentioned 3 For high accuracy do not block the left ai...

Page 8: ...Connect the test fixture or cable to LCUR LPOT HPOT and HCUR terminals on the instrument front panel Pay attention to the direction when inserting the test fixture to avoid abnormal measurement 1 7 Wa...

Page 9: ...r carefully 2 1 Introduction to Front Panel Figure 2 1 shows the front panel of TH2523 CAT 300VDC Max LCUR LPOT HPOT HCUR UNKNOWN PASS FAIL LOCK 7 4 1 9 6 3 8 5 2 COPY RESET TRIG SETUP DISP MENU 0 ENT...

Page 10: ...8 PASS FAIL Display The compare results will be displayed through PASS FAIL 9 USB Interface USB HOST Interface 10 RESET Key System reset key 11 TRIG Key In MAN manual mode pressing this key can trigg...

Page 11: ...can be conveniently constructed to realize auto test TH2523 will output bin comparator result signals and handshake signals by this interface meanwhile start up signal will also be sent to the instru...

Page 12: ...llowing zones as shown in figure 2 3 Figure 2 3 Display Zones 1 Main Menu Indicate the name of the currently displayed page 2 Function Test parameters can be modified in this zone 3 Measurement Result...

Page 13: ...ations of TH2523 keys are as follows Choose the desired page by menu keys DISP SETUP and soft keys Move the cursor to the destination zone by arrow keys The soft keys corresponding to the current curs...

Page 14: ...will be asked to input your password After inputting your password press ENTER to enter into the main menu page NOTE The series of products have enabled password protection and the default password is...

Page 15: ...ed in the screen as is shown in figure 3 1 Figure 3 1 Measurement Display The following measurement parameters can be set in this page Measurement Function FUNC R X Impedance Range Range R AUTO Voltag...

Page 16: ...ement Functions Measurable parameters on TH2523 are as follows Primary Parameters Z modulus of impedance R resistance L inductance V voltage C capacitance Secondary Parameters Q quality factor R ESR X...

Page 17: ...ng items will be displayed AUTO Set the RNG_R as AUTO HOLD Switch the RNG_R from AUTO to HOLD Since the RNG_R is set to HOLD the current measurement range will be locked and displayed in the RANGE zon...

Page 18: ...e 2 EXTER File 3 1 5 TOOLS Operation procedures are as below Move to the TOOL zone the following menus will be displayed in the soft key zone REL OFF Relative function on off When OFF is selected Rela...

Page 19: ...t be shorted correctly or it will result in relative data error and finally it will deviate from the accurate measurement result The correct short connection is shown as follows 3 2 BIN DISP Press the...

Page 20: ...good product sound GD good product sound OFF sound is off COMP ON means the compare function is open OFF means the compare function is not open LOAD BIN BIN1 9 The measurement result will be compared...

Page 21: ...y parameter trace will be displayed B secondary parameter trace will be displayed A B traces of primary and secondary parameters will be displayed simultaneously CURSOR Press this key and the measurem...

Page 22: ...HOLD 6V 60V TH2523 30V 300V TH2523A will be displayed in the soft key zone Press one of them voltage range will be automatically selected or locked The voltage range can be switched between 6V 60V and...

Page 23: ...d the secondary display value is the secondary measurement value minus the secondary reference value when is selected the secondary display value the secondary measurement value the secondary referenc...

Page 24: ...e rear panel trigger controlling measurement will also be done When Trig is set to the BUS mode the trigger measurement will be done through BUS 3 4 2 Delay Use the numerical keyboard to input the mea...

Page 25: ...OFF the bin comparison function of secondary parameter will be disabled NOM_A NOM_B Nominal value is only valid in mode Nominal value for primary and secondary parameters NOM_A and NOM_B You can input...

Page 26: ...through the numerical keyboard and press the corresponding soft key to confirm Interval Move the cursor to it and input the time interval between measurements through the numerical keyboard and press...

Page 27: ...he numerical keyboard and press the corresponding soft key to confirm When the secondary measurement result is larger than the datum set in B Stop1 or less than the datum set in B Stop 2 the instrumen...

Page 28: ...h the soft key There are two modes for comparison of high low value ABS resistance value inductance value and voltage value of Hi Lo and percentage of nominal value and Hi Lo NOTE the formula of perce...

Page 29: ...er results 3 7 5 Hi Lo Move the cursor to it and input the data through the numerical keyboard NOTE Upper and lower limit values should match Statis A and Statis B When the statistic mode is Statis A...

Page 30: ...pacity is insufficient 6 Hi num Be used to add up times that the measurement result exceeds the upper limit value 7 Lo num Be used to add up times that the measurement result is less than the lower li...

Page 31: ...peration Manual Ver1 2 26 will be displayed in the corresponding zone measurement times is the times set on the screen Pressing STAT STOP will stop counting TRIG Trigger the instrument to make measure...

Page 32: ...key area to enter into the SYSTEM SETUP page shown in the following figure 4 1 Figure 4 1 System Setup Page The following functions can be set in this page Key Sound Language Password Bus Mode Baud R...

Page 33: ...be displayed RS232C GPIB option USBTMC USBCDC 2 Press RS232C the RS232C interface is selected 3 Press GPIB the GPIB interface is selected NOTE GPIB is an option Only when it is installed this mode can...

Page 34: ...e For example 9 o clock 13 minute and 25 second a m on November 12 2010 will be shown as 10 11 12 09 13 25 Operations are as follows Touch the time zone to be modified the following items will be disp...

Page 35: ...a U disk data save external U disk CSV No Save measurement results to a U disk screen save external U disk GIF No Save the screen information to a U disk 4 2 2 Structure of File Folder File in a U Dis...

Page 36: ...emory exceeds 512M it is recommended to use FAT32 standard to format the disk 3 Before a U disk is connected to TH2523 you are recommended to save the data on it and Tonghui will not be liable for the...

Page 37: ...edited the following items will be displayed in soft key area LOAD Press LOAD if the file name that the cursor locates is not empty YES and NO will be displayed in the soft key zone When YES is select...

Page 38: ...Ver1 2 33 this file Copy to E Press Copy to E The instrument will copy the file the cursor locates or the selected file to a U disk Delete Press Delete if YES is selected the instrument will delete th...

Page 39: ...absolute value deviation ABS and percentage deviation between a measurement value and a programmable nominal value 5 1 4 Range Double range bin for resistance and DC voltage Mode AUTO MANU HOLD UP DO...

Page 40: ...5 digits maximum display number 31 000 4 digits maximum display number 3100 5 2 Test Signal 5 2 1 Frequency The test signal is a sinusoidal wave with the frequency of 1kHz 0 2Hz 5 2 2 Maximum Input Vo...

Page 41: ...short of the test cable is as follows Red Black White Blue 5 2 5 Basic Accuracy for Z L R X Measurement The following two tables display the basic accuracy of Z at SLOW MED and FAST SLOW MED acquirin...

Page 42: ...Fs 0 5 Rdg 0 01 Fs 3 3 3000 1m 10mA 10 30 33 000 10 m 1mA 10 300 330 00 100 m 100 A 10 3 k 3 5000 k 1 10 A 10 L C X accuracy using condition Dx D test value 0 1 R accuracy using condition Qx Q test va...

Page 43: ...0 V 100 V 0 05 Rdg 0 01 Fs 0 06 Rdg 0 01 Fs 60 V 65 000 V 1 mV FAST acquiring the basic accuracy at the temperature of 20 C 5 C the temperature coefficient plus 0 005 C Model TH2523 Display digits 4 R...

Page 44: ...accuracy at the temperature of 20 C 5 C the temperature coefficient plus 0 005 C Model TH2523A Display digits 4 Range Maximum display value Resolution Accuracy for 6 months Accuracy for one year 30 V...

Page 45: ...interface of TH2523 is also not strictly based on RS 232 standard but only uses the smallest subset of this standard The signals are listed in the following table Table 6 1 RS 232 Signal and Pin Conne...

Page 46: ...cate with PC or others intelligent equipments and meanwhile can make up automatic test system with the other test equipments Up to 15 equipments can be connected on a same bus TH2523 applies IEEE488 2...

Page 47: ...TH2523 Series Operation Manual Ver1 2 42 Figure 6 2 GPIB Connector Pin Structure Figure 6 3 Typical GPIB System Interconnection for Double piggyback Connector...

Page 48: ...le shows the TH2523 GPIB interface functions Code Capability Function SH1 Data Source Handshake Three wire Shake AH1 Acceptor Handshake Three wire Shake T5 Basic Talker Transmit instrument data L4 Bas...

Page 49: ...rate remotely DEVICE CLEAR SDC or DCL As soon as this command is received the selected device or all devices connecting with this interface will be cleared LOCAL GTL LOCAL returns to local control and...

Page 50: ...device or all listener devices It is necessary to assign TH2523 as a listener device firstly and then set the trigger mode as bus trigger mode Finally send trigger data 6 2 4 SCPI Standard Commands fo...

Page 51: ...ocols 6 3 1 System Configuration Connect USB interfaces on PC and TH2523 through a USB cable 6 3 2 Install the Driver When TH2523 is first connected to a PC through a USB cable the prompt information...

Page 52: ...ure 6 6 Procedure 2 of Installing USB Driver When the installation of driver is finished user can see usb test and measurement device in the device manager of PC as is shown in the following figure Fi...

Page 53: ...nd TH2523 to the USB interface through a USB cable Methods of installing USBCDC driver are the same with that of installing USBTMC driver When the installation of driver is finished user can see usb V...

Page 54: ...end signal of IEEE 488 bus 7 1 TH2523 Subsystem Commands DISPlay TRIGger COMParator APERture SYSTem FUNCtion FETCh Mass MEMory STATistical TRACe BINSETup 7 1 1 DISPlay Subsystem Commands DISPlay subsy...

Page 55: ...nvalid TSETup Set the display page to trace setup SYSTem Set the display page to system setup FLISt Set the display page to internal file list The STATe command is used to select whether to refresh th...

Page 56: ...2523 Series Operation Manual Ver1 2 51 The IMPedance command sets the function parameters The FUNCtion IMPedance query returns the current function parameters Command Syntax FUNCtion IMPedance functio...

Page 57: ...IMP RX Set function as R X Query syntax FUNCtion IMPedance Return Format function NL END The IMPedance RANGe command sets the instrument range The FUNCtion IMPedance RANGe query returns the current r...

Page 58: ...the range of DC voltage The FUNCtion VDC RANGe query returns the current DC voltage range parameter Command syntax FUNCtion VDC RANGe value Where value can be the positive integer from 0 1 0 300V TH25...

Page 59: ...Command syntax ON 1 FUNCtion SMONitor VAC OFF 0 Where Character 1 integer 49 means ON Character 0 integer 48 means OFF For example WrtCmd FUNC SMON VAC ON Set the voltage monitoring switch as ON Query...

Page 60: ...49 Sets the deviation mode of primary parameter Character 1 integer 50 Sets the deviation mode of secondary parameter For example WrtCmd FUNC DEV1 MODE ABS Query syntax FUNCtion DEV n MODE Return for...

Page 61: ...ry and secondary deviation nominal values For example WrtCmd FUNC DEV1 REF FILL The REL ON command controls the rel function OFF Command syntax FUNC REL ON Return format FUNC REL Return 1 integer 49 m...

Page 62: ...t measurement speed and average times used in measurement Command Syntax FAST APERture MEDium value SLOW Where FAST Fast 100meas sec MEDium Medium 50meas sec SLOW Slow 6 25meas sec value Average numbe...

Page 63: ...SOURce query returns the current mode of trigger source Command syntax INTernal TRIGger SOURce EXTernal BUS HOLD Where INTernal Be automatically triggered by the instrument and be the defaut setting...

Page 64: ...tion of 1mS MIN Set the delay time as 0S MAX Set the delay time as 60S For example WrtCmd TRIG DEL 5S Set the delay time as 5S the default unit is S Query syntax TRIGger DELay Return format NR3 NL END...

Page 65: ...re 1 integer 49 means ON 0 integer 48 means OFF For example WrtCmd COMP STAT ON Query syntax COMParator STATe Return format NR1 NL END The BEEper command sets the beeper of comparator function The COM...

Page 66: ...BIN Query syntax COMP CompMode Return value BIN bin mode COMP compare mode The LOADBinno is used to load the BIN number This command is only valid in compare mode The COMParator LOADBinno query retur...

Page 67: ...example WrtCmd BINSETup BinMode ABS Query syntax BINSETup BinMode Return value NR1 NL END NR1 is 0 absolute value mode 1 percentage mode The BINSETup COMPareA command set the primary parameter as ON...

Page 68: ...alA query returns the current value of the primary parameter Command syntax BINSETup NORmalA value Where value is floating point type ranged from 10k 10k For example WrtCmd BINSETup NORmalA 15 set the...

Page 69: ...N Return value NR3 NL END NR3 the upper and lower limit of the primary parameter The BINSETup BINB command set the upper and lower limit of the secondary parameter According to different modes the up...

Page 70: ...Manual Ver1 2 65 The STATIstics STATe command sets the statistical function of primary parameter or secondary parameter to ON or OFF Command syntax STATIstics STATe 1 2 A or B Where 1 integer 49 means...

Page 71: ...function to OFF Query syntax STATIstics STATUS Response NR1 NL END NR1 is 0 integer 48 means OFF 1 integer 49 means ON The STATIstics MODE command sets the mode of high and low limit in statistical f...

Page 72: ...sults Query syntax STATIstics MAXimum Response maximum NR3 serial number corresponding to the maximum NR1 For example Query STATI MAX Response 1 2450E 01 5 The STATIstics MINimum query returns the min...

Page 73: ...nction Command syntax STATIstics NORmalB value For example STATIstics NORmalB 50 Set the nominal value of secondary parameter to 50 the unit can be or V according to the selected measurement parameter...

Page 74: ...e Subsystem Commands The TRACe TOTAL value command is used to set and return the total time of scanning For example TRACe TOTAL 20 Query syntax TRACe TOTAL Response value NR2 The TRACe INTERval value...

Page 75: ...uery syntax TRACe BM Response max value NR3 min value NR3 The TRACe ASTOP1 OFF value TRACe ASTOP2 OFF value TRACe BSTOP1 OFF value TRACe BSTOP2 OFF value commands are used to set or return the stop po...

Page 76: ...BEEP OFF 0 ON 1 This command sets and returns Beep Tone For example SYST BEEP OFF Turn off beep tone Query syntax SYST BEEP Response NR1 NL END NR1 is 0 Turn off beep tone 1 Turn on beep tone 7 1 11 M...

Page 77: ...d MMEM LOAD STAT 1 The STORe STATe command saves the current settings into a file Command syntax MMEMory STOR STATe value name Where value The serial number of file ranging from 1 to 20 NR1 name File...

Page 78: ...H2523 s identification string Query syntax IDN Return foramt manufacturer model firmware NL END Where manufacturer Name of Manufacturer model Instrument Model firmware Firmware Version For example Wrt...

Page 79: ...ch byte of the status byte register are shown as follows Query syntax SRE Return format value NL END For example WrtCmd SRE The ESR query returns the contents of the standard event status register Que...

Page 80: ...set the OPC bit of the standard event status register when TH2523 finishes all parameter measurements Ever since all pending operations have been completed this command will inform the instrument to...

Page 81: ...TH2523 Series Operation Manual Ver1 2 76 Return format NL END Where 0 ASCII number 1 decimal number 49 For example WrtCmd OPC...

Page 82: ...lt output signal will be valid pin signal name description 1 9 BIN1 BIN9 In Compare mode the low bin number signal is effective 10 OUT In compare mode or bin mode if the compare result of BIN1 BIN9 ar...

Page 83: ...the low bin comparison result output signal is effective 25 KEYLOCK When this single is effective the keys in the front panel are locked Low signal is effective 26 NC Null pin 27 28 EXTV1 External po...

Page 84: ...TH2523 Operation Manual Ver1 0 79 EXTV1 Timing chart...

Page 85: ...nt the instrument should be handled with care and some precautions must be taken to resist moisture and water 9 4 Storage The instrument should be stored in an airy room where the environment temperat...

Page 86: ...tenance personnel Do not substitute the internal components unauthorized when maintaining In order to ensure the measurement accuracy the instrument must be measured and corrected after maintenance Yo...

Reviews: