![Tonghui Electronics TH1778A Operation Manual Download Page 14](http://html1.mh-extra.com/html/tonghui-electronics/th1778a/th1778a_operation-manual_1126457014.webp)
14
2)
According to the three parameters of current begin, current end and scan point to generate the list
automatically.
If current begin is I
b
, current end is I
e
, scan point is n, then the current step can be represented as I
s
=(I
e
-I
b
)
/(n-1). The current list can be shown as I
b
, I
b
+ I
s
, I
b
+ I
s
*2,
……, I
e
.
3)
Move the cursor to the desired place and make use of the numeris keyboard to input each value.
There are two methods to set the delay time:
1)
Move the cursor to the desired place and make use of the numeris keyboard to input each value.
2)
Move the cursor to the place where the delay time has been set already, and make use of the copy function to
copy the value to all the blanks.
3.3 System and Tool Settings
3.3.1 System Setting
3.3.1.1 Trig Mode
There are three trigger modes which are listed below:
Man: Make use of the start button on the front panel to start output current.
Ext: Make use of the pedal switch on the rear panel to start output current.
Bus: Make use of the RS232 interface on the rear panel to start output current.