As shown in the figure above, the setting interface includes parameter setting and list setting.
3.2.1 Parameter Setting
3.2.1.1 Current Output
Make use of the numeric keyboard to set the total current output value, the default unit is A.
The current setting range is listed below:
Slaver Amount
Set Lower Limit (A)
Set Upper Limit (A)
0
0
20
1
0
40
2
0
60
3
0
80
4
0
100
5
0
120
3.2.1.2 Response Frequency
Make use of the numeric keyboard to set the response frequency, the default unit is Hz.
The response frequency setting range is 0~2MHz.
3.2.1.3 Rise Time
Make use of the numeric keyboard to set the current rise time, the default unit is s.
3.2.1.4 Fall Time
Make use of the numeric keyboard to set the current fall time, the default unit is s.
3.2.1.5 Current Range
The instrument is designed to have three current ranges which are listed below:
Range
Current Range
1
0A~1A
2
1A~5A
3
5A~20A
The default is automatic range which means the instrument will automatically set to the appropriate range based
on the current setting value.