17
1. Alarm Level
The methane column density level at which an alarm threshold can be
set.
In the main menu select “1. Alarm level”
1.
The [ALARM LEVEL] screen shown on the right will
be displayed.
(The default setting value is 100 ppm
·
m.)
2.
Select a value with the
▲▼
buttons for the 1000’s
digit, and then press the Menu/Enter button.
❖
To leave the value as indicated just press the Menu/Enter button or the
Display/Back button as is.
3.
Likewise, select a value for the 100’s digit, 10’s digit, and 1’s digit.
4.
Once the setting for the 1’s digit is completed, the buzzer sounds twice
to indicate that the setting has been accepted.
5.
Your set value will be displayed on the numeric mode and graph mode
screen.
❖
The set value is not be reset even when the power is switched off.
❖
The alarm levels can be set up similarly for the other measuring units.
(See “7. UNIT” in this section.)
Set the [Alarm level] greater than the [Offset value]. If the [Alarm level] is
not set greater than the [Offset value], the alarm doesn't sound and the
detection lamp doesn't flash even when this product detects methane.
See [2. Offset Value] about how to set the [Offset value].
CAUTION