background image

   

34/35

 

Example 1) To delete data 001 to 200 
Set counter to 200 and press “MD” key. Confirm STT shows 1 and press “MD” to the next 
Example 2) To delete data 101 to 200 
Set counter to 200 and press “MD” key. Set STT number to 101 and press “MD” to the next 
Example 3) To delete all data 
Set counter to 999 and press “MD” key. Confirm STT shows 1 and press “MD” to the next 
 

iii. 

Press "MD" key and "C" key at the same time when the left display shows either "N", 
"MAX", "MIN" or "AVE" such as the following example displays. Then release both 
keys. 

 

     

 

 

 

 

 

iv. 

After deletion completed display back to the lower end counter and back to 
measurement mode 

・ 

The example display if there is no Hi/ Lo limit torque setting 

 

 

・ 

The example display if there are Hi/ Lo limit torque settings 

 

The selected data deletion completed 

Summary of Contents for CEM10N3X8D-G-BTA

Page 1: ...1 35 MODEL CEM3 G BTA Instruction Manual...

Page 2: ...le to transmit data wirelessly 1 2 Judgment features You can set the threshold for torque and angle the tool judge the result by your settings Trg Trigger torque TL Lower limit torque TH Higher limit...

Page 3: ...ning direction DN Tightening direction NG NG O Torque OK OK OK TL T TH L Torque LO NG NG LO NG T TL H Torque HI NG NG HI NG TH T OO Torque OK Angle OK OK OK TL T TH OK L w H OD Torque OK Angle DT NG N...

Page 4: ...UN mode when the memory counter displays 000 Without angle setting With angle setting 2 2 Maximum value display PEAK mode when the memory counter displays between 001 to 999 Without torque setting Wit...

Page 5: ...r cable C Clear key Clear saved measurement data On RUN mode to auto zero adjust POWER Power button When angle setting is on and turning on the power angle speed check will be automatically conducted...

Page 6: ...ce reaches on low limit torque value the tool give beep continuous and red LED When you set high and low limit torque under MODE T the left display shows T as below 3 2 Display mode RUN mode The count...

Page 7: ...nt When turning on the power angle speed check will be automatically conducted to make zero adjustment During zero adjustment the tool should be put still do not move otherwise the display may show Er...

Page 8: ...ction angle and direction of tightening these judge whether the measured result are within the range or not Under the Tightening mode MODE T when you set the high low torque and angle double tightenin...

Page 9: ...alarm When it exceeds 105 of the maximum measurable torque the value on the display and blinks alternatively and the buzzer does on 4 12 Over torque alarm Peak torque hold starting value Values in is...

Page 10: ...timer setting 0 5 ng NG output setting NG_MAN dCn Output format setting M3 ID Set both the angle setting and trigger torque setting to measure the angle i Set to Peak mode counter between 001 to 999 a...

Page 11: ...uts the torque and angle data When judgment is NG If you select NGAUTO in NG setting solid red LED inform NG result for one second and all measured data are automatically sent to external device If yo...

Page 12: ...deg An_h Higher limit angle setting 50 deg Ar Auto reset timer setting 0 5 ng NG output setting NG_MAN dCn Output format setting M3 ID Set both the angle setting and trigger torque setting to measure...

Page 13: ...ds continuously the left display shows the NG judgment result and measured angle repeated alternatively each 1 second About NG judgment result refer the 1 2 Judgment features Press MEM key to save and...

Page 14: ...ion specifications Bluetooth version Ver 3 0 Communication method AFH Modulation system GFSK Radio output power 4 dBm Transmission power class Class 2 Profile SPP Communication distance Approx 10 m Ac...

Page 15: ...Input command Action Discription Note AT008 Change the torque unit Send numbers instead of N m 06 kgf cm 02 kgf m 03 lbf in 08 lbf ft 09 AT037 Set the higher and lower limit torque values Send torque...

Page 16: ...N3X15D G BTA Send a command the AT037 090 0 080 0CRLF Example 3 Set to 600 Nm for higher 500 Nm for lower limit for CEM850N3X32D G BTA Send a command the AT037 0600 0500 CRLF When set to the 15 degree...

Page 17: ...e Lower limit angle An_L 0 0 999 Enable when set on the trigger torque More than double tightening detection angle Higher limit angle An_H 0 0 999 Enable when set on the trigger torque More than doubl...

Page 18: ...ure setting When turn on the angle feature At reset power on or leaving setting mode after change the angle feature setting from OFF to ON an angle speed check will be automatically conducted At RUN m...

Page 19: ...C key to go back to RUN mode measurement mode iii Unit of torque Default N m Select the unit of torque measurement from N m kgf cm kgf m lbf in or lbf ft Select the unit using upper or down arrow keys...

Page 20: ...the double tightening detection angle Hi Lo limit angle settings are skipped and changed to 0 Then the display goes to tightening direction setting Set the number using down arrow key and move the dig...

Page 21: ...r limit angle from trigger torque If the value is set less than double tightening detection angle or lower limit angle the display shows an error SETERR Set the number using down arrow key and move th...

Page 22: ...is declined Press MEM key to output data or C key to clear NGAUTO Auto output Auto reset feature is applied even if though the judgment result is NG Select the mode using upper or down arrow keys Pres...

Page 23: ...per or down arrow keys Press MEM key to save setting and go to next Press MD key to skip setting and go to next Press C key to go back to RUN mode measurement mode xv Communication mode Default BT Sel...

Page 24: ...next Press MD key to skip setting and go to next Press C key to go back to RUN mode measurement mode xvii Data length Default 8bit Select the data length from 7 bit or 8 bit Set to 8 bit when use 584...

Page 25: ...ment mode xix Communication format Default M 3 Select the communication format It affects the data strings for communication Refer the 7 3 Communication format for details dCn M 3 dCn M3 ID When chang...

Page 26: ...r the time and date settings Select the DFT Y or DFT N using upper or down arrow keys When display shows DFT Y and press MEM key the tool is factory reset to save setting and go to next Press MEM key...

Page 27: ...e measurement mode Minute setting Set minute using upper or down arrow keys Press MEM key to save setting and go to next Press MD key to skip setting and go to next Press C key to go back to RUN mode...

Page 28: ...arrow keys Press MEM key to save setting and go to next Press MD key to skip setting and go to next Press C key to go back to RUN mode measurement mode Date setting Set year using upper or down arrow...

Page 29: ...u need using keys then press MD key to the next Example 1 To output data 001 to 200 Set counter to 200 and press MD key Confirm STT shows 1 and press MD to the next Example 2 To output data 101 to 200...

Page 30: ...th right display turns off and left display shows numbers from lower to upper end during batch data output NOTE Press C key to cancel data output Other keys won t work during data output Batch data ou...

Page 31: ...ext Example 1 To process data 001 to 200 Set counter to 200 and press MD key Confirm STT shows 1 and press MD to the next Example 2 To process data 101 to 200 Set counter to 200 and press MD key Set S...

Page 32: ...s minimum value of the selected data range Press MD key to proceed or press C key to go back to measurement mode vi The display shows minimum value of the selected data range Press MD or C key to go b...

Page 33: ...e is no Hi Lo limit torque setting The example display if there are Hi Lo limit torque settings 9 3 2 Delete the selected range of data i Set the counter number to the upper end that you delete using...

Page 34: ...to 999 and press MD key Confirm STT shows 1 and press MD to the next iii Press MD key and C key at the same time when the left display shows either N MAX MIN or AVE such as the following example disp...

Page 35: ...er once and turn it on without touching any other keys If error message disappears then it operates normally If error message remains on the display contact TOHNICHI or your nearest distributo to ask...

Reviews: