viii. Lower limit angle (Default: 0)
Set the lower limit angle from trigger torque
※
If the value is set less than double tightening detection angle the display shows an
error "SETERR".
Set the number using down arrow key and move the digit using upper arrow key. Press
“MEM” key to save setting and go to next. Press “MD” key to skip setting and go to next.
Press “C” key to go back to RUN mode (measurement mode).
ix. Higher limit angle (Default: 0)
Set the higher limit angle from trigger torque
※
If the value is set less than double tightening detection angle or lower limit angle,
the display shows an error "SETERR".
Set the number using down arrow key and move the digit using upper arrow key. Press
“MEM” key to save setting and go to next. Press “MD” key to skip setting and go to next.
Press “C” key to go back to RUN mode (measurement mode).
x.
Tightening direction (Default: CW)
Set the tightening direction from CW, CCW or both
Select the direction using upper or down arrow keys. Press “MEM” key to save setting
and go to next. Press “MD” key to skip setting and go to next. Press “C” key to go back
to RUN mode (measurement mode).