![TNI softFlow Instructions For Use Manual Download Page 37](http://html.mh-extra.com/html/tni/softflow/softflow_instructions-for-use-manual_1122046037.webp)
37
Instructions for Use
TNI
soft
Flow
Instructions for Use
TNI
soft
Flow
Device parameters
Technical data
Medical product class (93/42/EWG)
IIa
Safety class, electrically
II
Alarm signal sound pressure
> 45 dB(A)
Safety type
IP21
(Protected from touch by fingers and objects
>12 mm and vertically dripping water)
Applied part (applicator)
BF
Electrical safety
According to EN 60601-1
UL 60601-1
CSA C22.2/No 60601-1
Electromagnetic compatibility
According to EN 60601-1-2
Operating voltage (nominal voltage)
100-240 V
~
, 50-60 Hz
Maximum power system
300 VA
Maximum power applicator heating
20 VA
Device dimensions
Width
320 mm
Depth
320 mm
Height
210 mm
Weight (without humidifier, without water)
5.6 kg
Humidification chamber auto-fill
max. 144 ml
Sterile water bag or bottle
< 1000 ml
Water tank humidifier homecare
max. 650 ml
Applicators softFlow mode
Changing cycle applicator clinic series
Changing cycle applicator homecare series
≤ 360 therapy hours; single-patient use
≤ 720 therapy hours; single-patient use
Safety level (applied part)
BF
Tube length nasal application
1.8 m
Tube length tracheal application
2,33 m
Max. temperature of air leaving device
43°C
Applicators junior mode
Changing cycle NeoFlow nasal cannula
Changing cycle InnoTube softFlow
7 days; single use
30 days; single use
Safety level (applied part)
BF
Tube length
160 cm
Max. temperature of air leaving device
43 °C