21
Instructions for Use
TNI
soft
Flow
Instructions for Use
TNI
soft
Flow
2.2.6 Tracheal Application
Turn on the TNI softFlow device before attaching an applicator. Attach the applicator to the patient’s tra-
cheostomy interface according to the following sequence of images.
Connect the tracheal interface to
the patient connection according
to manufacturer specifications.
Connect the applicator tube
adapter to the matching coun-
terpart of the tracheal interface.
Check the connection between
the applicator and patient inter-
face for strong hold.
NOTE
• The applicator tube connector has an inner diameter of 22mm. Only use suitable interfaces
when selecting the patient interface compliant with ISO 5367 or ISO 80601-2-74.
• This accessory can only be used with the system operated in softFlow mode.
•
Humidification system output during tracheal application is >33 mg / l.
WARNING
•
Oxygen admixture is limited to 20 l / min when using the Applicator Tracheal-Interface! This
must be taken into consideration when making therapy decisions and adjustments.
• During tracheal application the humidity is automatically set to 37°C DP and cannot be
changed as long as the tracheal applicator is in use.
•
Recommended ambient temperature during tracheal application is 20° - 28°C which differs
from the recommended range of 18° - 28°C for nasal application.