Instructions for Use
TNI
soft
Flow
Instructions for Use
TNI
soft
Flow
191
I
Check application
hose
Warning: Flow obstruction detected. Check if
application hose was kinked.
201
II
Air flow too hot
Therapy air temperature is too high. Please
check environmental conditions and refer to
the product specifications.
251
I
Ambient tempe-
rature off limits
Ambient temperature is out of permitted ran-
ge. Please refer to the product specifications.
252
I
Ambient humidi-
ty off limits
Ambient humidity is out of permitted range.
Please refer to the product specifications.
254
I
Sensor defective
Ambient temperature / humidity sensor is
defective.*
255
I
Dew point not
reachable
Set dew point cannot be reached. Please refer
to the product specifications and check the
correct set-up of the system components.
301
II
Heating plate
gets too hot
Hardware error.*
302
II
Heating plate
defective
Heating plate electronic is not working pro-
perly.*
351
I
Please refill water Please refill the humidification chamber with
water.
352
I
Heating plate
defective
Heating plate is not working properly.*
353
I
Sensor defective
Temperature sensor of humidifier is defective.*
354
I
Heating plate
defective
Heating plate is not working properly.*
355
I
Sensor defective
System failure.*
401
II
Blower defective
Blower is blocked.*
402
II
Blower gets too
hot
Blower is overheated. Please refer to the pro-
duct specifications and check the air flow.
403
II
Blower sensor
defective
Blower temperature sensor is defective.*
404
II
Fan defective
Fan blower is defective.*
501
II
-
Display defective; an acoustic alarm signal is
given.*
502
II
System failure
Sensor errors detected on system startup.*
503
II
Calibration error
Restart device with closed oxygen supply.
Slowly reopen oxygen after the device has
reached set air flow completely.*
601
II
Sensor defective
Temperature sensor of applicator is defective;
please replace by a new applicator.
605
II
Air flow too hot
Therapy air temperature is too high. Please
check environmental conditions and refer to
the product specifications.