Return: 

 

This saves the options and returns you to the main screen. 

 

STATISTICS 
 
 

The 49-86 and 49-87 models have a statistics feature that allows the user to display the average, 

standard deviation, maximum, and minimum values of the readings stored in the instrument’s 

temporary memory (this memory is erased when the power to the instrument is turned off). 

 

1.

 

To access the main statistics screen from the main (measurement) screen, press the 

UNITS

 button. 

2.

 

The main statistics screen will display the current reading, the average of the stored readings, the 
standards deviation of the stored readings and the maximum and minimum of the stored readings.   

3.

 

Press the 

UNITS

 button again to list the readings 

a.

 

Navigate this screen by pressing the ZERO key to highlight ↑ , ↓ , DELETE, or RETURN

 

b.

 

Press the UNITS key to act on the highlighted option 

i.

 

Pressing UNITS when the ↑ is highlighted will scroll the active reading selection 

box up once. 

ii.

 

Pressing UNITS when the 

↓ is highlighted will scroll the active reading selection 

box down once. 

iii.

 

Pressing UNITS when the DELETE is highlighted will delete the active reading 
in the selection box. 

iv.

 

Pressing UNITS when the RETURN is highlighted will return to the main 
statistics screen. 

c.

 

Press and hold the UNITS key, then press the ZERO key to delete all readings. User 
confirmation is required for this. 

4.

 

Press and hold the

 UNITS 

key, then the 

ZERO

 key to return to the Main (measurement) screen. 

 

SERIAL DATA OUTPUT 
 

These Testing Machine Inc. Micrometers (Thickness Gauges) output a string through a mini-USB 
connector on the back panel of the instrument.  
 
The USB host (PC) will need to be looking for serial data on one of its COM ports. The number of the 
COM port will vary from host to host. 
 
These are the settings needed to capture the serial data - 
 

Baud rate: 9600 

 

Data: 8 bits 

 

Parity: None 

 

Stop Bit(s): 1 

 

Flow Control: None 

 
Here are some serial string examples- 
 

x

 

Initial string - "CC5.1.1.13;" 

o

 

This is an identifier string used by our GraphMaster product. You can ignore it if you are 
not using GraphMaster. It will only show up after powering on. 

 

x

 

After each cessation of downward motion, like the spindle coming to rest on a sample, the 
machine will output a serial data string in the following format. 
 
TM V00.01.02 49-87-00-000112345-67  0000000004.370 
 

o

 

"TM V00.01.02 " - A software version number 
 

digital-micrometer-ld3581-49-86-m44.pdf   15

digital-micrometer-ld3581-49-86-m44.pdf   15

15-6-2020   14:37:59

15-6-2020   14:37:59

Summary of Contents for 49-85

Page 1: ...NES INC The Finest Test Equipment for all Industries 49 85 86 87 MICROMETERS PLEASE READ BEFORE OPERATING UNIT digital micrometer ld3581 49 86 m44 pdf 1 digital micrometer ld3581 49 86 m44 pdf 1 15 6...

Page 2: ...y through Friday from 8 30 AM to 5 00 PM EST TMI also offers on site Preventative Maintenance and Calibration service Please take the time to fill out the postage paid Warranty Card completely and ret...

Page 3: ...ting of 600v 10 Do not remove any panels from the tester Never push objects of any kind into any openings in the instrument s covers Failure to heed this warning could result in fire or electric shock...

Page 4: ...described below are required In case of eye contact Flush the eyes with plenty of clean water for at least 15 minutes immediately without rubbing Take a medical treatment If appropriate procedures ar...

Page 5: ...CALIBRATION GAUGE 7 FEATURES 7 NAME AND FUNCTION OF PARTS 8 SETUP 12 OPERATION 12 INTRODUCTION 16 PARALLELISM CHECK ADJUSTMENT 16 STATISTICS 15 SERIAL DATA OUTPUT 15 MAINTENANCE 18 digital micrometer...

Page 6: ...well as a movable pressure foot is made from lapped stainless steel APPLICATION The TMI Micrometer is designed to measure the thickness or caliper of paper sheets films and foils Almost any material...

Page 7: ...unted for CALIBRATON GAUGE To check the calibration properly the user of the micrometer should have the necessary gauge blocks on hand The gauge blocks can be ordered separately from TMI See Section 3...

Page 8: ...00 15 Standard Model 2 pieces of the afore mentioned part number make this weight 6 Lower Anvil A testing surface that is optically flat The specimen is placed on the anvil s surface and when the pres...

Page 9: ...the AC power inlet of the instrument before opening the fuse holder Do not use fuses with a higher current rating then the original as this can cause damage to the instrument The use of repaired fuse...

Page 10: ...49 85 86 87 10 Rev E Figure 1 digital micrometer ld3581 49 86 m44 pdf 10 digital micrometer ld3581 49 86 m44 pdf 10 15 6 2020 14 37 59 15 6 2020 14 37 59...

Page 11: ...49 85 86 87 11 Rev E Figure 2 11 10 13 12 digital micrometer ld3581 49 86 m44 pdf 11 digital micrometer ld3581 49 86 m44 pdf 11 15 6 2020 14 37 59 15 6 2020 14 37 59...

Page 12: ...er plug needs to be adapted to the local power requirements this must be done by a qualified electrical engineer The plug must be equipped with a safety earth terminal Please Note Power cord wire colo...

Page 13: ...hat are kinked bent or creased should not be used since this may affect the reading of the sample 9 Measure the thickness by placing the specimen between the automatically cycling pressure foot and th...

Page 14: ...g is within the tolerance band around zero Zero Threshold This sets a small range around a reading of 0 00 to be reported as 0 00 For example if the machine is set to a Zero Threshold of 2 m then any...

Page 15: ...election box iv Pressing UNITS when the RETURN is highlighted will return to the main statistics screen c Press and hold the UNITS key then press the ZERO key to delete all readings User confirmation...

Page 16: ...e steel gauge blocks of different sizes See Section 3 Specifications These gauge blocks are checked against standards that are traceable to the NIST The part number for the TMI gauge blocks is as foll...

Page 17: ...49 85 86 87 17 Rev E Figure 1 digital micrometer ld3581 49 86 m44 pdf 17 digital micrometer ld3581 49 86 m44 pdf 17 15 6 2020 14 38 00 15 6 2020 14 38 00...

Page 18: ...ted recheck the parallelism For instruments 49 86 00 0003 and 49 87 00 0003 perform three sets of readings adjusting the screws each iteration in order to establish a baseline parallelism Only perform...

Page 19: ...ons or order acknowledgements provided All materials delivered services provided and work performed by TMI are subject to the Terms and Conditions listed on the linked document To view and or download...

Page 20: ...digital micrometer ld3581 49 86 m44 pdf 20 digital micrometer ld3581 49 86 m44 pdf 20 15 6 2020 14 38 00 15 6 2020 14 38 00...

Reviews: