Introduction
This operation manual is designed for familiarization with the basic operation of the device, particu-
lars of its design, storage rules, and operational procedure.
Intended purpose and application
The device
UV-Radiometer “ТКА – UV»
(hereinafter referred to as the device) is designed for
measuring irradiance (in mW/m
2
) created:
in the spectral range
UV - А
(315 … 400) nm by UV sources,
in the spectral range
UV - В
(280 … 315) nm by UV sources,
in the spectral range
UV - С
(220 … 280) nm by UV sources.
The sphere of application of the device: sanitary and technical surveillance in residential and indus-
trial rooms, certification of workplaces, and other spheres of activity.
Basic technical characteristics
Wight, kg (not above) with power module
2.0
Operation parameters
Ambient temperature,
о
С:
– normal operation conditions
– operation temperature range
20
±
5
0…40
Relative air humidity at ambient temperature25°С, %, not above
95
Atmospheric pressure, kPa
80…110
Mean time between failures with confidence probability р = 0.8, h, not below
2000
Irradiance measurement
•
Measurement range, mW/m
2
1 … 60 000
•
Limit of the permitted basic relative error
of irradiance measurements, %
±
10,0
Continuous operation from battery, hours (not below)
from the mains, hours (not below)
2,0
8,0
Overall dimensions, mm (not over):
— photometer head
— laptop ASUS Eee PC 701 (closed)
185х70х55
225x165x35
It is prohibited to measure high irradiance over a long time!
It is recommended to use a heat-resistance lens cover supplied with the device.
Generated by Foxit PDF Creator © Foxit Software