8-2
Table 8-2. Error Code Descriptions
Recommended Troubleshooting
Sequence
Error No.
Brief description
Checked
During
Use?
RFGen
Bd
Power
Supply
Display
Bd
Test
Resistor Remarks
001
CRC check error
1
002
RAM test error
1
003
CRC check error
YES 1
004
Watchdog error
1
005
Local I²C BUS errors (EEPROM)
YES 1
006
Controller I²C BUS error
YES 1 2
007
µController defective
YES 1
008 Software inconsistency
YES 1
009
Heat sink MP1
YES 1 2
Note 1
010
Heat sink MP2
YES 1 2
Note 1
011
Case temperature
YES 1 2
Note 1
013
int. A/D converter, reference voltage and
analog multiplexer
YES 1 2
014
Ground
YES 1 2
015
High power supply 0 V test
1
2
016 High power supply watchdog test/
1
2
017
Power supply Enable test
1
2
018
Power down test
1
2
019
High power supply error
U test
1
2
020
High power supply error
I test
1
2
023
MPU1 self-test error
YES
Note 2
026
Power level 2 error
1
3 2
027
Power level 3 error
1
3 2
028
Power level 4 error
1
3 2
029
Power level 5 error
1
3 2
030
Power level 6 error
1
3 2
032
RF enable error
1
033
Current level 0 error
1
3 2
034
Current level 1 error
1
3 2
035 Watchdog reset
YES 1
036 Excess output: power
YES 1 3 2
037 Faulty variable contents
YES 1
038 Mutual time monitoring of MPUs
YES 1
039
Power supply voltage exceeds tolerances
YES 1 3 2
040
Power supply current exceeds tolerances
YES 1 3 2
041
Oscillator frequency exceeds tolerances
YES 1
042
Pump voltage exceeds tolerances
YES 2 3
Note 3
043
Pump current exceeds tolerances
YES 2 3
Note 3
044
Power down capacitor test
1
2
045 Power failure occurred
YES
046 Command error
YES 1
Summary of Contents for Aquamantys System