_______________________________________________________________________________
6-6
Figure 6-4.
Initiating the Priming Sequence
The “Start Prime” button activates and deactivates the timed priming cycle.
Pressing the button a second time will immediately stop the priming cycle.
Pressing the button a third time will reset the timer and restart the priming
cycle from the beginning.
Test of visual indicators
During the power-up self-test of the Aquamantys Pump Generator,
verify that all visual indicators illuminate.
Test of alarm tone and volume control function
Using manual procedure described in the RF Leakage Current section
above, activate the bipolar output of the pump generator. Verify that
there is an audible activation tone. Also verify that the volume of
that tone can be adjusted (but not turned down to an inaudible
range) with the volume control on the rear panel of the pump
generator.
Power cord inspection
Inspect the power cord for any signs of exposed wires, cracks,
frayed edges, or connector damage. Replace the power cord with an
appropriate hospital grade replacement if any of these conditions or
other evidence of damage exists.
Replacement power cords may be ordered from TissueLink Medical
Customer Service by calling 866.777.9400 (+1.603.742.1515
outside U.S.).
Fuse check
Warnings:
Shock Hazard.
Turn off and unplug pump generator prior to accessing the
fuse holder.
Check the rating of the fuses in the line filter on the rear of the
Aquamantys Pump Generator for correct ratings. To do so, unplug
the power cord from the generator. Using a flat bladed screwdriver,
eject the fuse holder out of the line filter, remove the microfuses and
check for correct rating: