3-2
Activation Tone
Frequency (nominal)
940 Hz
Alarm Tone
Frequency (nominal)
349, 415, 524, 698 Hz
Leakage Currents
See IEC test record
Electromagnetic Compatibility
The Aquamantys Pump Generator is intended for use in the
electromagnetic environment specified below. The customer or user
of the unit should assure that it is used in such an environment.
Immunity Test
IEC (60)601-1-2:2001
Test Level
Conducted emission
DIN EN 55011, FCC Part 18,
class B, consumer class
150 kHz - 30 MHz:
Radiated emission
DIN EN 55011, FCC Part 18,
class B, non consumer class
30 MHz - 1 GHz
Electrostatic discharge
DIN EN 61000-4-2
± 6 kV Contact discharge
± 8 kV Air discharge
Immunity to electromagnetic fields
DIN EN 61000-4-3
10 V/m
80 - 2500 MHz
Immunity to conducted fast transients
DIN EN 61000-4-4
Burst:
± 2 kV power mains
± 1 kV signallines
Immunity to conducted slow transients
DIN EN 61000-4-5
Surge 1.2/50µs:
± 2 kV
unsym
/± 1 kV
sym
power mains
Immunity to conducted disturbances
induced by RF-fields
DIN EN 61000-4-6
10 V
rms
150 kHz - 80 MHz
power mains / signallines
Voltage dips, short interruptions
DIN EN 61000-4-11
Complies
Harmonic current emission
DIN EN 61000-3-2, class A
Complies
Voltage fluctuation and flicker
DIN EN 61000-3-3
Complies
LEDs
All LEDs inside the Aquamantys are CLASS 1 LED PRODUCT
according to EN60825-1.
Input Power
The nominal mains voltage is factory selected. Refer to the rear
panel markings for correct mains voltage.