PUF+ CONNECTIONS
The descriptions of the various connections to the PUF+ control unit are described
below.
“Motor”
The “Motor” power connection supplies power to the motor being controlled by the
PUF+. Both Brush-Type and Brushless-Type motor are supported, but only one may be used
at a time. The PUF+ will come configured to match the motor type ordered.
THE MOTOR
USED WITH THE PUF+ MUST MATCH THE AC POWER BEING SUPPLIED TO
THE PUF+. WHILE THE PUF+ CONTROLS THE MOTOR SPEED THROUGH THE
EMBEDDED ELECTRONICS, IT CAN NOT CHANGE THE OPERATING VOLTAGE
OR THE FREQUENCY OF THE INCOMING POWER THAT IS SUPPLIED TO THE
MOTOR.
“Power”
The “Power” connection supplies power to the PUF+. It should be connected to the
AC power line. Operation at both 120 Volts AC and 240 Volts AC are supported. The
system is designed to operate at either 50 or 60 Hz without the need to change the
configuration of the unit. However, a jumper inside the PUF+ determines the voltage on
which the unit operates.
DO NOT APPLY 240-VOLT POWER TO THE PUF+ WHILE
THE UNIT IS CONFIGURED FOR 120-VOLT OPERATION. THIS WILL DAMAGE
THE UNIT. APPLYING 120-VOLT POWER TO THE PUF+ WHILE THE UNITS IS
CONFIGURED FOR 240-VOLT OPERATION SHOULD NOT DAMAGE THE UNIT,
BUT IS NOT RECOMMENDED OR SUPPORTED
.