8. G value 

In the weighing model, press and hold

tare

key, and then press

print

key

it shows 

UF-1. Press 

tare

  key for eight times

it shows UF-9. Press

print

key

it shows 

-00-

and then shows 9.79423

G value of the manufacturing place

. And then press 

print

  key to adjust the G value of the using place, the first number in the left 

begins to twinkle, and press

net/gross

key to choose the number. Press

ZERO

key to move cursor to left or press

tare

key to move cursor to right. Press

print

key to confirm, and press

units

key to back to weighing model. 

Calibration: 

1. Weight and zero calibration 

 

In the weighing model, press and hold

print

  key, and then press 

ZERO

key, 

it shows ECF-1. Press 

print

  key, it shows CAL Z to begin zero calibration. Press 

print

key, it shows the weight which was calibrated last time. Use numeric key to 

change the weight, and then put the required weight on the plate. And then press

print

key to finish calibration.

 

2.

 Zero calibration

 

In the weighing model, press and hold 

print

key, and then press 

ZERO

key, 

it shows ECF-1. And then press 

net/gross

key, it shows ECF-2. Press 

print

key, 

it shows CAL Z. Press 

print

key to calibrate zero and back to weighing model.

 

3. Weight calibration

 

In the weighing model, press and hold 

print

  key, and then press

ZERO

  key, 

it shows ECF-1. And then press

net/gross

  key twice, it shows ECF-3. Press 

print

 

key, it shows the weight which was calibrated last time in the Weighing Window. Use 
numeric key to change the weight, and then put the required weight on the plate. And 
then press   

print

  key to finish calibration.   

 

Error Messages 

 

Err  H   

  Initial zero too high (over 10% of max. cap.) 

 

Err  L         Initial zero too low (less than 10% of max. cap.)   

 

Err  N   

  Unstable internal value 

           

hhhhhh      Overload, over the max. ca 9d 

 

------       Total price exceeded 999999 

 

Remark          
         

0      0     0 

         A    B    C 
A: Buzzer on: 0= Stable not required 1= Stable required 
B: LCD indicator and RELAY on: 0= Stable not required 1= Stable required 
C: Buzzer beeps when: 0= Buzzer off      1= OK      2= LO and HI

 