![Tiger Optics HALO M7000 Series Instruction Manual Download Page 33](http://html.mh-extra.com/html/tiger-optics/halo-m7000-series/halo-m7000-series_instruction-manual_1115521033.webp)
HALO Users Manual
Revision G
25
Tool Bar
Measurement results
Status Bar
Figure 4-2 Main Display
The
Main Display
provides easy access to all indicators and controls. The Main Display has five areas:
the tool bar, the measurement results, system status, alarm status, and the status bar.
#
The
Tool Bar
buttons are used to either select the operating mode, or to bring up auxiliary windows.
#
The
Measurement Results
area displays the measured contaminant concentration.
#
The
Mode
status indicates the current mode selection: Service, Tune, or Measure.
#
The
Laser
status indicates the status of the laser source. The display will show either OK or ALARM
#
The
Ringdown
status indicates the status of the ringdown acquisition. The display will show either
OK or ALARM
#
The
Alarm Status
areas indicated if the current value exceeds the user selectable alarm points. The
display will show either OK or ALARM
#
The
Status bar
indicates the current setting of various operating parameters. From left to right, these
are: laser current, laser temperature, ringdown threshold level, the ring-down time of the gas being
sampled (Tau Measured), the off-peak ringdown time of the system (Tau Zero), and the approximate